Table 1.
Types of commissioning and QA tests
Type | Description |
---|---|
A | Qualitative assessment of registered image, this typical involve using visualization technique such as examining the difference image between the registered image and the reference image. |
B | Qualitative assessment of DVF. |
C | Qualitative assessment of contours on registered images such as overlaying anatomical structures defined on one image, and these can be overlaid on another image. |
D | Quantitative assessment of contours on registered images using metrics such as DICE coefficient and Hausdorff distance. |
E | Quantitative assessment of landmark alignment using TRE. |
F | Assessment of Jacobian determinant |
G | Assessment of other DVF metrics. |
H | Consistency and transitivity measurements, these techniques include performing the registration in both directions to ensure that the registration is inverse consistent. |
I | Quantitative assessment using digital phantoms, these can also be useful to quantitatively assess the accuracy of image registration. |
J | End to end tests using physical phantoms, this ensures that all the different part of the radiotherapy system workflow works accurately. |
DVF, deformation vector field; QA, quality assurance; TRE, target registration error.