Skip to main content
. 2021 Apr 29;94(1122):20210001. doi: 10.1259/bjr.20210001

Table 2.

Summary of the different applications that were used clinically, including the most common clinical sites, type of commissioning data used and frequency of ongoing QA

Application Number of centres (% of 20 clinical users) Most common anatomical sites Length of time in clinical use Types of commissioning data Ongoing QA frequency
Propagate contours from one scan to another 19 (95%) Head & neck (17)
Prostate (12)
Lung, rectum, gynae (eight each)
<1 year (4)
1–3 years (11)
>3 years (4)
Patient (13)
Digital phantom (4)
Physical phantom (3)
Patient-specific (13)
None (6)
Registering pre-treatment multimodality imaging (e.g., target delineation with PET or MRI) 10 (50%) Head & neck (7)
Brain, liver, prostate, gynae (two each)
<1 year (3)
1–3 years (4)
>3 years (3)
Patient (7)
Digital phantom (3)
Physical phantom (1)
Patient-specific (6)
Monthly (1)
Annual (1)
None (3)
Deform planning CT to daily images 8 (40%) Head & Neck (6)
Lung, liver, prostate, Gynae (four each)
<1 year (1)
1–3 years (7)
>3 years (0)
Patient (7)
Digital phantom (2)
Physical phantom (1)
Patient-specific (6)
None (2)
Dose propagation 7 (35%) Head & Neck (6), lung (4), prostate (4) <1 year (1)
1–3 years (6)
>3 years (0)
Patient (5)
Digital phantom (2)
Physical phantom (1)
Patient-specific (6)
None (1)

PET, positron emission tomography; QA, quality assurance.