Table 2.

Sensitivity of lateral flow devices: Approach 1 (known positives). Estimates based on analysis of 268 individuals self-reporting previous PCR confirmed infection. CI = confidence interval, UI = uncertainty interval based on percentiles from Monte Carlo simulation, TPs = true positives, FNs = false negatives, “Probability best” = the proportion of Monte Carlo simulations in which the test had the highest sensitivity. Note: these AbC-19TM results have been published previously (21) and are reproduced here for comparative purposes.

Lateral flow immunoassay
AbC-19TM Orient Gene 1 band Orient Gene 2 bands SureScreen 1 band SureScreen 2 bands Biomerica 1 band Biomerica 2 bands

TPs 248 252 252 238 252 253 255
FNs 20 16 16 30 16 15 13
Sensitivity
(95% CI)
92.5% (88.8, 95.1) 94.0% (90.5, 96.3) 94.0% (90.5, 96.3) 88.8% (84.5, 92.0) 94.0% (90.5, 96.3) 94.4% (91.0, 96.6) 95.1% (91.9, 97.1)
Rank
(95% UI)
6
(2, 6)
4
(1, 6)
4
(1, 6)
7
(7, 7)
3
(1, 6)
3
(1, 6)
1
(1, 5)
Probability
best
0.01 0.08 0.08 0.00 0.14 0.04 0.64
False negatives by days since between symptom onset and blood sample:
Asymptomatic (n=12) 5 3 3 5 4 4 4
8-21 days (n=5) 0 0 0 0 0 0 0
22-35 days (n=20) 1 1 1 2 1 1 1
36-70 days (n=142) 6 4 4 12 5 2 1
≥71 days (n=89) 8 8 8 11 6 8 7