Skip to main content
. Author manuscript; available in PMC: 2022 May 14.
Published in final edited form as: Physiol Meas. 2021 May 14;42(4):10.1088/1361-6579/abf886. doi: 10.1088/1361-6579/abf886

Table 7.

Calibration lines data for poP bare bone (t = 0 mm) and poP bone with overlapping POM, resin and wax ST phantoms of four different thickness values. The number in the round parentheses is the uncertainty in the last one or two significant figures of the corresponding value.

ST Thickness t (mm) Pb Lα Pb Lβ
Slope(s)(keV μg−1g) y-intercept(keV) R2 Slope(s)(keV μg−1g) y-intercept(keV) R2
0.00 0.073(2) 0.09(7) 0.996 0.073(5) −0.1(2) 0.975
POM 1.10(5) 0.035(3) 0.14(11) 0.973 0.042(5) −0.1(2) 0.932
1.89(5) 0.015(3) 0.33(12) 0.839 0.023(2) −0.03(8) 0.962
3.23(3) 0.0076(13) 0.25(6) 0.868 0.016(2) 0.20(13) 0.914
4.36(3) 0.006(3) 0.29(11) 0.494 0.0119(13) 0.03(5) 0.945
Resin 1.10(5) 0.048(2) −0.14(9) 0.989 0.052(5) −0.2(2) 0.958
2.18(5) 0.036(4) 0.03(18) 0.932 0.030(2) 0.14(10) 0.968
3.05(3) 0.013(2) 0.12(10) 0.857 0.020(4) −0.02(18) 0.811
4.19(3) 0.006(2) 0.11(10) 0.583 0.009(2) 0.14(10) 0.738
Wax 1.47(8) 0.046(6) 0.0(2) 0.923 0.046(2) −0.15(9) 0.989
2.04(8) 0.0355(16) 0.11(7) 0.989 0.025(6) 0.1(3) 0.760
3.48(8) 0.026(3) 0.30(13) 0.933 0.033(4) 0.02(15) 0.945
4.01(5) 0.021(3) 0.16(13) 0.898 0.022(4) 0.05(16) 0.872