Skip to main content
. Author manuscript; available in PMC: 2022 May 14.
Published in final edited form as: Physiol Meas. 2021 May 14;42(4):10.1088/1361-6579/abf886. doi: 10.1088/1361-6579/abf886

Table 8.

Pb detection limit (DL) values for bare bone and ST phantom calibration line data.

ST Thickness t (mm) Pb Lα Pb Lβ
σ0 (counts keV) DL (μg g−1) σ0 (counts keV) DL (μg g−1)
0.00 0.068 2.76(8) 0.179 7.3(5)
POM 1.10(5) 0.105 9.0(7) 0.281 20(2)
1.89(5) 0.123 24(5) 0.297 39(3)
3.23(3) 0.090 35(6) 0.187 35(3)
4.36(3) 0.079 40(18) 0.138 35(3)
Resin 1.10(5) 0.088 5.5(3) 0.281 16.2(1.5)
2.18(5) 0.174 15(2) 0.265 26(2)
3.05(3) 0.114 27(5) 0.250 38(8)
4.19(3) 0.092 45(17) 0.161 54(14)
Wax 1.47(8) 0.108 7.0(9) 0.302 19.6(9)
2.04(8) 0.123 10.4(5) 0.286 34(9)
3.48(8) 0.149 17(2) 0.328 30(3)
4.01(5) 0.107 15(2) 0.262 35(6)