Table 3. Reliability metrics for hip force measurements using a custom rig.

Session 1 Session 2 ICC (95% CI) SEM (95% CI) SEM MDC (95% CI) MDC
(N) (N) (N) % (N) %
Abduction (n = 20) 120 ± 26 124 ± 29 0.77 (0.51–0.90) 13 (10–19) 11 31 (26–35) 25
Adduction (n = 20) 158 ± 33 159 ± 28 0.89 (0.71–0.96) 10 (8–16) 7 24 (23–26) 15
Extension (n = 18) 287 ± 69 270 ± 66 0.90 (0.73–0.97) 21 (16–33) 8 49 (33–66) 18
Flexion (n = 17) 234 ± 42 239 ± 51 0.80 (0.53–0.92) 21 (16–32) 9 49 (44–53) 21
Internal rotation (n = 20) 171 ± 43 169 ± 39 0.81 (0.57–0.92) 18 (14–27) 11 42 (40–43) 25
External rotation (n = 20) 125 ± 25 130 ± 29 0.77 (0.48–0.91) 13 (10–19) 10 30 (25–35) 23
E+AB (n = 20) 208 ± 45 207 ± 52 0.86 (0.66–0.95) 19 (14–29) 10 45 (40–50) 23
E+ER (n = 20) 211 ± 54 196 ± 54 0.86 (0.68–0.94) 18 (13–27) 9 42 (41–42) 20
E+AB+ER (n = 20) 201 ± 55 197 ± 54 0.88 (0.70–0.95) 20 (15–29) 10 47 (32–62) 23

Notes.

Values are mean ± standard deviation unless otherwise indicated.

ICC
intraclass correlation coefficient
CI
confidence interval
SEM
standard error of measurement
MDC
minimal detectable change
N
newton
E
extension
AB
abduction
ER
external rotation