. 2021 Apr 1;35(6):e23776. doi: 10.1002/jcla.23776

Diagnostic value of loop‐mediated isothermal amplification assay for hand, foot, and mouth disease

Ying‐Zhou Chen 1,2, Zhi‐Qing Zhan 1,3, Li‐Quan Zhou 1,2, Min‐Shan Chen 1,2, Xun‐Jie Cao 1,3, Ya‐Ping Li 1,4, Xu‐Guang Guo 1,3,5,6,✉
PMCID: PMC8183915  PMID: 33792998

Abstract

Background

Nowadays, hand, foot, and mouth disease (HFMD) has a significant negative impact on children's health, especially in the Asia‐Pacific region. Loop‐mediated isothermal amplification assay (LAMP) is a highly efficient and convenient novel tool. However, its diagnostic accuracy for HFMD is still not clear. Therefore, we conducted a meta‐analysis in order to evaluate the potential of LAMP assay for the diagnosis of HFMD, in which the reference standard was polymerase chain reaction (PCR).

Methods

A protocol was predetermined (CRD42020212882) in PROSPERO. We retrieved seven databases including PubMed for relevant studies published before October 2020. Articles were included if they compared the diagnostic efficiency of LAMP with PCR for HFMD through detecting clinical samples which was more than 15. Statistical analysis was performed by STATA 15.1 software. Risk of bias and applicability were assessed using Quality Assessment of Diagnostic Accuracy Studies. No funding was used for the study.

Results

A total of 18 retrospective studies including 2495 samples from China were finally included. Reference standards of them included RT‐PCR and non‐RT‐PCR. The merged sensitivity and specificity with 95% confidence interval (95% CI) were 1.00 (0.97–1.00) and 0.97 (0.88–0.99), respectively. The pooled PLR, NLR, and DOR with 95% CI were 11.17 (5.91–21.11), 0.05 (0.03–0.09), and 538.12 (183.17–1580.83), respectively. The AUC of SROC was 1.00 (95% CI: 0.99–1.00).

Conclusion

In conclusion, our research revealed high sensitivity and specificity of LAMP in diagnosing HFMD. However, more high‐quality research is required to prove this conclusion.

Keywords: diagnosis, hand foot and mouth disease, loop‐mediated isothermal amplification, meta‐analysis, polymerase chain reaction


Our research revealed high sensitivity and specificity of LAMP in diagnosing HFMD. The AUC of SROC was 1.00 (0.99–1.00), indicating the high diagnostic accuracy of LAMP for HFMD.

graphic file with name JCLA-35-e23776-g004.jpg

1. INTRODUCTION

Hand, foot, and mouth disease (HFMD) is a common infectious disease in children caused by Enterovirus A, in which the most generally reported genotypes are Enterovirus A71 (EV‐A71) and Coxsackievirus A6, A10, and A16 (CVA6, CV‐A10, and CV‐A16). 1 It is common in preschoolers. In China, children below the age of 3 are most susceptible to HFMD. 2 HFMD is characterized by macular papules and herpes on the palms, soles, oral mucosa, and buttocks, with or without fever. 3 It is prone to central nervous system damage and severe cases of brainstem encephalitis, leading to neurogenic pulmonary edema, pulmonary hemorrhage, and even death. 4 Timely and correct diagnosis is of great significance for its treatment.

The traditional laboratory diagnostic methods of HFMD are mainly isolation culture and serological detection. However, because of the need for a particular culture medium, time‐consuming and complex experimental conditions, and comparatively low detection rate leading to late diagnosis, isolation culture has not been routinely used in HFMD diagnosis. The practical value of serology in the diagnosis of early HFMD infection is limited since serological diagnosis is time‐consuming. 5 It is usually used as a retrospective diagnosis and epidemiological investigation to provide the basis for the formulation of prevention and control measures. 6 In recent years, with the rapid development of molecular biotechnology, real‐time fluorescent quantitative polymerase chain reaction (PCR) has become a routine method for laboratory diagnosis of HFMD. 7 , 8 However, a lot of money is required in the early stages to meet the needs of the environment, personnel, and instruments, so it is difficult to detect this in a primary laboratory. With the shortcoming of requiring costly precise equipment, PCR may not be a good method to diagnose HFMD in the basic clinical place, especially for developing countries. 9 Therefore, it is necessary to find a rapid, economical, simple, sensitive, and specific tool to diagnose HFMD.

Loop‐mediated isothermal amplification (LAMP) is a newly emerged nucleic acid amplification method, which relies upon DNA polymerase with strand replacement ability and two pairs of primers, which can identify six regions of the target sequence. 10 The whole detection reaction can be completed only at 65℃ for 1–2 h, and the results can be judged by the naked eye. Compared with other detection methods, this method is simple to operate and the equipment used is not expensive. 11 The most prominent feature of LAMP is that it allows instant diagnosis. 12 It has a widespread application in the diagnosis of infectious diseases such as malaria and trypanosomiasis. 13 Moreover, it has been regarded as a potential alternative to PCR in the laboratory. 14 However, the diagnostic accuracy of LAMP for HFMD diagnosis remains unclear. Hence, we performed a pooled‐analysis to evaluate the overall efficacy of LAMP in diagnosing HFMD. PCR was served as the reference standard.

2. MATERIALS AND METHODS

2.1. Protocol and registration

A protocol was predetermined in PROSPERO, and the registration number was CRD42020212882, which could be accessed at https://www.crd.york.ac.uk/prospero/display_record.php?ID=CRD42020212882. This research was conducted following Preferred Reporting Items for a Systematic Review and Meta‐analysis of Diagnostic Test Accuracy Studies (PRISMA‐DTA). 15

2.2. Search strategy

We identified all the studies published before October 2020 by systematically searching Embase, PubMed, Cochrane Library, Web of Science, Wan Fang Data, SinoMed, and the Chinese National Knowledge Infrastructure (CNKI) databases using the following strategy as follows: (“Hand, Foot, Mouth Disease” OR “Hand, Foot and Mouth Disease” OR “hand foot and mouth disease” OR “hand foot mouth disease”) AND (“LAMP” OR “isothermal amplification loop‐mediated” OR “loop‐mediated isothermal amplification” OR “loop mediated isothermal amplification”). Then, we inspected the bibliographies of all the publications to complement the retrieval.

2.3. Study selection

Two review authors carefully reviewed all the retrieved articles and selected the articles according to the inclusion and exclusion criteria established in advance. If they had different opinions on some articles, they discussed them with a third review author until they were in agreement.

2.4. Inclusion and exclusion criteria

Articles were included in the analysis if (1) clinical samples of human were analyzed; (2) they compared the diagnostic accuracy of LAMP with PCR for HFMD; (3) PCR was served as the reference standard; (4) the sample size was more than 15; (5) the generated data sufficed to construct two‐by‐two tables including true positive (TP), false positive (FP), false negative (FN), and true negative (TN) for working out the sensitivity, specificity, and likelihood ratios.

Criteria for excluding the studies were as follows: (1) Duplicate publications; (2) Abstracts, case reports, letters, reviews, and editorials.

2.5. Data extraction

The following information was extracted individually by two review authors from the articles selected for inclusion: Publication information, sample size, specimen type, gold standard, and virus detected type. Data for 2·2 tables (TP, FP, FN, and TN) were extracted. When inconsistencies were encountered, the two review authors resolved them by reaching an agreement with a third one.

2.6. Quality assessment

Relying on the Quality Assessment of Diagnostic Accuracy Studies (QUADAS‐2) guidelines, 16 two review authors assessed the risk of bias from 4 aspects: Patient selection, indicator testing, gold‐standard method, and timing and flow. They also considered the applicability concerns from the first three aspects. The final figure was made by Review Manager 5.3.

2.7. Statistical analysis

Stata 15.1 software was employed to perform the statistical analysis. We calculated the proportion of heterogeneity likely due to threshold effect and drew the summary receiver operating characteristic (SROC) curve to tell whether there was a threshold effect. Next, we calculated the merged diagnostic odds ratio (DOR), sensitivity, specificity, positive likelihood ratio (PLR), and the negative likelihood ratio (NLR) of the included studies and corresponding 95% confidence intervals (95% CI). We used Cochran's Q test and I 2 test to evaluate the heterogeneity derived from the non‐threshold effect of the eligible studies. If the p value of Q test >0.05 or I 2 < 50%, a fixed‐effects model was applied. Otherwise, we adopted a random‐effect model. Meta‐regression analysis was also conducted to validate whether heterogeneity was caused by different types of samples, reference standards, or the kinds of virus detected. The area under the SROC curve (AUC) was calculated. We plotted Deeks’ funnel plot to detect publication bias.

2.8. Subgroup meta‐analysis

Subgroup analysis was carried out depending on whether the standard reference was reverse transcription polymerase chain reaction (RT‐PCR) and on the types of samples, respectively.

3. RESULTS

3.1. Literature searching result

A total of 422 studies were identified from which 281 duplicates were discarded. After screening their titles and abstracts, 119 articles were excluded, and 22 articles were left to browse the full‐text. Finally, three articles were removed because of the lack of gold standard or due to insufficient data to construct a 2×2 table. Another article was discarded since the index test was combined with additional techniques. The remaining 18 studies 9 , 33 which were eligible were included to conduct pooled‐analysis. The flow diagram is shown in Figure S1.

3.2. Characteristics of the eligible studies

All these eligible articles were retrospective studies with a total of 2495 samples from China. These 18 articles were published between 2011 and 2017. The sample size ranged from 20 to 678. Funding Sources of these articles did not include related reagent makers. Seven 9 , 18 , 20 , 24 , 27 , 28 , 32 of them included clinical samples from suspected patients with HFMD while eleven of them included clinical samples from confirmed patients. All the patients had symptoms associated with HFMD. Five 17 , 18 , 20 , 21 , 25 of them generated data from only fecal samples while three 29 , 31 , 32 of them reported data from only pharyngeal swabs. The rest used a variety of samples. Besides, eight 9 , 18 , 19 , 22 , 24 , 25 , 27 , 33 of them detected only Human enterovirus 71 (EV71), and two 21 , 23 of them detected Coxsackievirus A16 (CA16). The remaining 8 examinations detected both EV71 and CA16. The baseline characteristics are summarized in Table 1.

TABLE 1.

Characteristics of the included studies

First author Year country study design reference standard Sample size sample type virus type TP FP FN TN
Zhao 2011 China retrospective PCR 60 Fecal samples EV71&CA16 34 4 0 22
Xia 2011 China retrospective RT‐PCR 108 Fecal samples EV71&CA16 87 14 0 7
Jiang 2011 China retrospective rRT‐PCR 40 Fecal samples EV71 26 0 2 12
Zong 2011 China retrospective rRT‐PCR 33 Fecal samples CA16 7 0 0 26
Geng 2011 China retrospective RT‐PCR 58 Fecal samples Pharyngeal swabs EV71 20 13 0 25
Shi 2011 China retrospective PCR 122

Fecal samples

Pharyngeal swabs

Vesicular fluids

EV71 56 2 0 65
Nie1 2012 China retrospective qRT‐PCR 145 Nasopharyngeal swab EV71 112 0 17 16
Yan 2012 China retrospective RT‐PCR 68 Fecal samples EV71 50 13 0 5
He1 2012 China retrospective rRT‐PCR 33

Fecal samples

Throat swards

rectal swards

vesicular fluid

EV71 24 0 0 9
He2 2012 China retrospective rRT‐PCR 33

Fecal samples

Throat swards

rectal swards

vesicular fluid

CA16 7 0 0 26
Nie2 2013 China retrospective rPCR 515

Fecal samples

Pharyngeal swabs

Liver swab CSF

EV71&CA16 336 2 0 177
Wang 2014 China retrospective RT‐PCR 36

Fecal samples

Pharyngeal swabs

EV71 23 5 0 8
Ding 2014 China retrospective qRT‐PCR 261 unclear EV71&CA16 169 0 17 75
Dai 2014 China retrospective qRT‐PCR 93 Pharyngeal swabs EV71&CA16 80 2 0 11
Guan 2015 China retrospective rRT‐PCR 92 unclear CVA16&other enteroviruses 9 0 0 83
Sun 2015 China retrospective RT‐PCR 20 Pharyngeal swabs EV71&CA16 11 1 3 5
Gao 2016 China retrospective rPCR 678 Pharyngeal swabs EV71&CA16 251 1 0 427
Li 2017 China retrospective RT‐PCR 100 Fecal samples Pharyngeal swabs EV71 60 3 2 35

Abbreviations: CA16, Coxsackievirus A16; CSF, cerebrospinal fluid; EV71, Human enterovirus 71; FN, false negative; FP, false positive; PCR, polymerase chain reaction; qRT‐PCR, real‐time quantitative reverse transcription polymerase chain reaction; rRT‐PCR, real‐time reverse transcription polymerase chain reaction; RT‐PCR, reverse transcription polymerase chain reaction; TN, true negative; TP, true positive.

3.3. Quality assessment

Most studies had a low risk of bias in most domains. And they all had low concern for applicability. However, 11 (61.1%) of the studies were considered to have a high risk of bias in the patient selection domain due to the inclusion of participants with confirmed diagnoses. Risk of bias of another two studies 17 , 20 was assessed as high in the index test domain because they did not pre‐specify a threshold. Reference standard results of Xia's study 17 were not interpreted without knowledge of the results of the index test. Therefore, this study was considered as high risk in the reference standard domain. The results of independent studies and the overall results are shown in Figures 1 and 2, respectively.

FIGURE 1.

FIGURE 1

Quality assesstion graph of the included studies

FIGURE 2.

FIGURE 2

Quality assesstion summary of the included studies

3.4. Threshold effect

“Shoulder‐arm” distribution of scatter plots did not exist in the SCOR curve (Figure 3). The proportion of heterogeneity likely due to threshold effect was 0.17. There was probably no threshold effect in the included studies.

FIGURE 3.

FIGURE 3

Summary receiver operating characteristic (SROC) curves of HFMD diagnosed by LAMP

3.5. Heterogeneity analysis of the non‐threshold effect

The non‐threshold effect was explored mainly using DOR. As shown in Figure S2, I 2 = 56.2% (I 2 > 50%) and p = 0.002 (p < 0.05), which suggested that the heterogeneity caused by non‐threshold effect existed. The random‐effect model was adopted to conduct the meta‐analysis. Nevertheless, high heterogeneity was also discovered through other indicators: sensitivity (I 2 = 94.0%, p < 0.05), specificity (I 2 = 95.9%, p < 0.05), PLR (I 2 = 88.9%, p < 0.05), and NLR (I 2 = 52.9%, p < 0.05).

3.6. Merge analysis results

To assess the diagnostic efficacy of LAMP assay using a random‐effect model, 18 sets of fourfold table data from 18 studies were merged. The pooled DOR was 538.12 (95% CI: 183.17–1580.83). The merged PLR, NLR, sensitivity, specificity, and corresponding 95%CI were 11.17 (5.91–21.11), 0.05 (0.03–0.09), 1.00 (0.97–1.00), and 0.97 (0.88–0.99). The results were presented in Figures S2–S4, and 4.

FIGURE 4.

FIGURE 4

Forest plots for the pooled sensitivity and specificity of LAMP

3.7. SROC curve

The calculated AUC was 1.00 (95% CI: 0.99–1.00) as presented in Figure 3, indicating the high diagnostic accuracy of LAMP for HFMD.

3.8. Meta‐regression analysis and subgroup analysis

It could be observed in Figure 5 that only the reference standard variable had statistical significance for sensitivity (p < 0.001). The results of the subgroup analysis were shown in Table 2.

FIGURE 5.

FIGURE 5

Meta‐regression results

TABLE 2.

Subgroup analysis results

Variables Subgroup Number of studies Sensitivity (95% CI) I 2 Specificity (95% CI) I 2
PCR technique RT‐PCR 14 0.98 (0.92–1.00) 89.0% 0.98 (0.80–0.99) 94.1%
Non‐RT‐PCR 4 1.00 (0.98–1.00) 0.0% 0.98 (0.92–1.00) 91.1%
Sample type Fecal samples 5 1.00(0.81–1.00) 91.2% 0.82(0.37–0.97) 92.1%
Pharyngeal swabs 3 0.99(0.81–1.00) 87.2% 0.97(0.71–1.00) 77.1%
Fecal samples and Pharyngeal swabs 3 0.99(0.92–1.00) 4.5% 0.85(0.63–0.96) 87.5%

3.9. Publication bias

As shown in Figure 6, significant publication bias was discovered in Deeks’ funnel plot (p = 0.00).

FIGURE 6.

FIGURE 6

Deeks’ funnel plot asymmetry test to assess publication bias

4. DISCUSSION

Hand, foot, and mouth disease poses a heavy disease burden mainly in Asia and the Pacific Rim, especially among children under the age of 5. In some severe cases, it can result in death due to fatal cardiopulmonary and neurological complications. 34 Thus, it is of great significance to master rapid and effective diagnostic methods for the treatment of HFMD. Our study concentrated on investigating the diagnostic value of the LAMP assay for HFMD.

After merging the results of 18 articles, we gained the merged sensitivity, specificity, PLR, NLR, and DOR, which were 1.00, 0.97, 11.17, 0.05, and 538.12, respectively. DOR was so high that it was obvious that LAMP had a high diagnostic value for HFMD. Nevertheless, the following items also proved the point. First, the SROC curve was far away from the middle diagonal and lay close to the upper left corner. At the same time, its AUC was 1.0. Second, NLR < 0.1 and PLR > 10, suggesting that LAMP was reliable in judging whether a person had HFMD. Third, high sensitivity and specificity suggested that the possibility of misdiagnosis and missed diagnosis were both low. Nevertheless, it was found that higher sensitivity and specificity existed in the group that used both pharyngeal swabs and fecal samples than the one that used just one of them, which indicated that various samples might improve the diagnosis value. Additionally, compared with the group that took RT‐PCR as a reference standard, the non‐RT‐PCR group had higher sensitivity and specificity, which indicated that using non‐RT‐PCR as a reference standard might overstate the diagnosis accuracy.

Typical diagnostic methods for HFMD are virus isolation and serological detection. However, they are too time‐consuming to meet the requirement of treatment of patients in hospital. Recently, PCR‐based method is generally used for laboratory diagnosis of HFMD. However, it relies on operator skills and expensive instruments, which is difficult to be wisely used especially in developing countries. 35 , 36 Compared to the methods discussed above, LAMP has some outstanding features, such as a wide detection range and simpler equipment. The efficiency of LAMP is almost impervious to the presence of non‐target genomic DNA in the reaction, which is appropriate to the improvement of diagnostic systems. 10 , 37 Moreover, this meta‐analysis reveals that the sensitivity and specificity of LAMP for HFMD are high. According to this meta‐analysis, it can be inferred that LAMP has the potential to be an alternative diagnostic tool for HFMD in the laboratory. It can be applied early in various fields and contribute to control the spread of infection. Nevertheless, it can also be well popularized and equipped in economically underdeveloped areas, which is helpful for the rapid control of HFMD epidemic in these areas. 37

Regarding heterogeneity, it was apparent that the heterogeneity caused by the effect threshold probably did not exist. However, the high heterogeneity derived from the non‐effect threshold could not be overlooked. In the random effects model, I 2 values of pooled DOR, sensitivity, specificity, PLR, and NLR were 56.2%, 94.0%, 95.9%, 88.9%, and 52.9% (p < 0.05), which exceeded 50%. Based on the results of meta‐regression, it could be observed that only the reference standard variable had statistical significance for sensitivity (p < 0.001), indicating that the sensitivity results obtained by studies using the gold standard of RT‐PCR and non‐RT‐PCR had statistically significant differences. Therefore, it was believed that the reference standard was a noticeable source of heterogeneity. Subsequently, a subgroup analysis was also performed. The heterogeneity of sensitivity in the non‐RT‐PCR subgroup decreased to 0, while another subgroup still exceeded 50% (decreased to 89.0%). It indicated that there were some extra potential sources of heterogeneity, such as the severity of the disease, different operators or conditions of the test, age or gender of the patients, and sample size.

Moreover, Deeks’ funnel plot was made to test publication bias. It showed that significant publication bias existed in our study (p < 0.05), which probably resulted from some unpublished negative results. That is to say, we were supposed to draw a conclusion based on the merged results with prudence.

Our study still had some limitations. First, the articles we included all originated in China, so the lack of data from other countries might impact the results. Second, some studies were considered to have a high risk of bias in certain domains, which might increase the risk of bias. Studies that did not refrain from a case‐control design might change the diagnostic accuracy. 38 In Xia's and Zhao's study, they did not pre‐specify a threshold for the index test, which might result in measurement bias. In Xia's study, interpretation of reference standard results with knowledge of the index test results might also increase the measurement bias. Third, the sources of high heterogeneity were not figured out completely. Besides, it was unclear whether unpublished negative results affected the existing results significantly.

In conclusion, according to the present combined results, LAMP has high sensitivity and specificity for the diagnosis of HFMD. That is to say, it has a high diagnostic value for HFMD. With the characteristic of high efficiency and convenience, LAMP may become the main laboratory diagnostic tool for HFMD in the future. However, more high‐quality research is required to prove this conclusion.

CONFLICT OF INTEREST

The authors declare that there are no competing interests associated with the manuscript.

Supporting information

Fig S1

Fig S2

Fig S3

Fig S4

DATA AVAILABILITY STATEMENT

Not applicable.

REFERENCES

  • 1. Knowles NJ, Hovi T, Hyypiä T, et al. Family Picornaviridae. In: King AMQ, Adams MJ, Carstens EB, eds. Virus Taxonomy: Ninth Report of the International Committee on Taxonomy of Viruses. London: Academic Press; 2012:855‐880. [Google Scholar]
  • 2. Xing WJ, Liao QH, Viboud C, et al. Epidemiological characteristics of hand‐foot‐and‐mouth disease in China, 2008–2012. Lancet Infect Dis. 2014;14(4):308‐318. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Guidelines for the diagnosis and treatment of hand foot and mouth disease (2018 version). Chin J Viral Dis. 2018;8(05):347‐352. [Google Scholar]
  • 4. Wang SM, Liu CC, Tseng HW, et al. Clinical spectrum of enterovirus 71 infection in children in southern Taiwan, with an emphasis on neurological complications. Clin Infect Dis. 1999;29(1):184‐190. [DOI] [PubMed] [Google Scholar]
  • 5. Lin TL, Li YS, Huang CW, et al. Rapid and highly sensitive coxsackievirus an indirect immunofluorescence assay typing kit for enterovirus serotyping. J Clin Microbiol. 2008;46(2):785‐788. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Wu KJ, Chen GT. Current status and progress in clinical laboratory diagnosis of hand, foot and mouth disease. Med Recapitulate. 2013;19(018):3332‐3335. [Google Scholar]
  • 7. Puenpa J, Suwannakarn K, Chansaenroj J, Chansaenroj J, Vongpunsawad S, Poovorawan Y. Development of single‐step multiplex real‐time RT‐PCR assays for rapid diagnosis of enterovirus 71, coxsackievirus A6, and A16 in patients with hand, foot, and mouth disease. J Virol Methods. 2017;248:92‐99. [DOI] [PubMed] [Google Scholar]
  • 8. Niu PH, Qi SX, Yu BZ, et al. Development of a highly sensitive real‐time nested RT‐PCR assay in a single closed tube for detection of enterovirus 71 in hand, foot, and mouth disease. Arch Virol. 2016;161(11):3003‐3010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Shi WF, Li K, Ji Y, Jiang Q, Shi M, Mi Z. Development and evaluation of reverse transcription‐loop‐mediated isothermal amplification assay for rapid detection of enterovirus 71. BMC Infect Dis. 2011;11(1):197. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Notomi T, Okayama H, Masubuchi H, et al. Loop‐mediated isothermal amplification of DNA. Nucleic Acids Res. 2000;28(12):E63. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Dou M, Sanjay ST, Dominguez DC, Zhan S, Li X. A paper/polymer hybrid CD‐like microfluidic SpinChip integrated with DNA‐functionalized graphene oxide nanosensors for multiplex qLAMP detection. Chem Commun. 2017;53(79):10886‐10889. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Wong YP, Othman S, Lau YL, Radu S, Chee HY. Loop‐mediated isothermal amplification (LAMP): a versatile technique for detection of micro‐organisms. J Appl Microbiol. 2018;124(3):626‐643. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Cai ZH, Dai YY, Huang LY, , Zhang WS, Guo XG. Diagnosis of mycoplasma pneumoniae by loop‐mediated isothermal amplification: systematic review and meta‐analysis. BioMed Central. 2019;19(1):173. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Li Y, Fan P, Zhou S, Zhang L. Loop‐mediated isothermal amplification (LAMP): a novel rapid detection platform for pathogens. Microb Pathog. 2017;107:54‐61. [DOI] [PubMed] [Google Scholar]
  • 15. McInnes MDF, Moher D, Thombs BD, et al. Preferred reporting items for a systematic review and meta‐analysis of diagnostic test accuracy studies: the PRISMA‐DTA statement. JAMA. 2018;319(4):388‐396. [DOI] [PubMed] [Google Scholar]
  • 16. Whiting PF, Rutjes AW, Westwood ME, et al. QUADAS‐2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med. 2011;155(8):529‐536. [DOI] [PubMed] [Google Scholar]
  • 17. Xia JF, Yan XF, Yu H, Qu D, Long JE. Simple and rapid detection of human enterovirus 71 by reverse‐transcription and loop‐mediated isothermal amplification: cryopreservation affected the detection ability. Diagn Micr Infec Dis. 2011;71(3):244‐251. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Jiang T, Liu J, Deng YQ, et al. Development and evaluation of a reverse transcription‐loop‐mediated isothermal amplification assay for rapid detection of enterovirus 71. J Clin Microbiol. 2011;49(3):870‐874. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Geng YZ, Yao WQ, Guo JQ, et al. Establishment and application of RT‐LAMP technique in detection of enterovirus 71 gene. Chinese J Zoonoses. 2011;27(03):176‐179. [Google Scholar]
  • 20. Zhao GP. Development and application of loop‐mediated isothermal amplification methods for enterovirus 71 and coxsackievirus A16 detection [dissertation]. Hebei: Hebei Medical University. 2011. [Google Scholar]
  • 21. Zong WP. Molecular epidemiological features of coxsackievirus A16 in Shenzhen and establishment of RT‐LAMP [dissertation]. edn. GuangDong: Southern Medical University; 2011. [Google Scholar]
  • 22. He YQ, Zong WP, Wang XH, et al. Detection of human Enterovirus 71 reverse transcription loop‐mediated isothermal amplification (RT‐LAMP). Lett Appl Microbiol. 2012;54(3):233‐239. [DOI] [PubMed] [Google Scholar]
  • 23. He YQ, Zong WP, Wang XH, et al. Detection of Coxsackievirus A16 by reverse transcription loop‐mediated isothermal amplification. Future Virol. 2012;7(5):519‐525. [Google Scholar]
  • 24. Nie K, Qi SX, Zhang Y, et al. Evaluation of a direct reverse transcription loop‐mediated isothermal amplification method without RNA extraction for the detection of human enterovirus 71 subgenotype C4 in nasopharyngeal swab specimens. PLoS ONE. 2012;7(12):e52486. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Yan XF. Epidemiological characteristics of children with hand‐foot and mouth disease in Shanghai and the early laboratory diagnose of EV71 Infection [dissertation]. edn. ShangHai: Fudan University; 2012. [Google Scholar]
  • 26. Nie K. Study of pathogen of hand, foot, and mouth disease(HFMD) by isothermal amplification [dissertation]. HeBei: Hebei Medical University; 2013. [Google Scholar]
  • 27. Wang D, Wang X, Geng Y, An C. Detection of enterovirus 71 gene from clinical specimens by reverse‐transcription loop‐mediated isothermal amplification. Indian J Med Microbi. 2014;32(2):124‐129. [DOI] [PubMed] [Google Scholar]
  • 28. Ding X, Nie K, Shi L, et al. Improved detection limit in rapid detection of human enterovirus 71 and coxsackievirus A16 by a novel reverse transcription‐isothermal multiple‐self‐matching‐initiated amplification assay. J Clin Microbiol. 2014;52(6):1862‐1870. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Dai Y, Liu FT, He S, et al. Rapid and economical detection of pathogens of hand‐foot‐and‐mouth disease by reverse transcription loop‐mediate amplification. Med Innov China. 2014;11(25):68‐71. [Google Scholar]
  • 30. Guan L, Xu ST, Nie K, et al. Colorimetric detection of coxsackievirus A6 by reverse transcription loop mediated isothermal amplification. Chin J Prev Med. 2015;49(11):1009‐1013. [PubMed] [Google Scholar]
  • 31. Sun XZ, Cheng XD, Zheng NY, et al. Isolation and identification of virus related with HFMD based on LAMP. Guangzhou Chem Ind. 2015;43(04):131‐134. [Google Scholar]
  • 32. Gao J. Application and comparison of RT‐LAMP and real time PCR in detection of hand, foot and mouth disease. World Latest Med Inform. 2016;16(18):176‐177. [Google Scholar]
  • 33. Li S, Zhang RS, Chen JF. Establishment and preliminary application of RT‐LAMP rapid detection method of entervirus 71. Int J Lab Med. 2017;38(04):470‐472. [Google Scholar]
  • 34. Duan XX, Chen ZH, Li XZ, Yuan P, Long L. Virus shedding in patients with hand, foot and mouth disease induced by EV71, CA16 or CA6: systematic review and meta‐analysis. Pediatr Infect Dis J. 2021;40(4):289–294. [DOI] [PubMed] [Google Scholar]
  • 35. Jiang B, Zhang J, You X, et al. Diagnosis of hand, foot, and mouth disease caused by EV71 and other enteroviruses by a one‐step, single tube, duplex RT‐PCR. J Med Virol. 2012;84(11):1803‐1808. [DOI] [PubMed] [Google Scholar]
  • 36. Singh S, Chow VT, Phoon MC, Chan KP, Poh CL. Direct detection of enterovirus 71 (EV71) in clinical specimens from a hand, foot, and mouth disease outbreak in Singapore by reverse transcription‐PCR with universal enterovirus and EV71‐specific primers. J Clin Microbiol. 2002;40(8):2823‐2827. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37. Mori Y, Notomi T. Loop‐mediated isothermal amplification (LAMP): a rapid, accurate, and cost‐effective diagnostic method for infectious diseases. J Infect Chemother. 2009;15(2):62‐69. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38. Rutjes AW, Reitsma JB, Vandenbroucke JP, Glas AS, Bossuyt PM. Case‐control and two‐gate designs in diagnostic accuracy studies. Clin Chem. 2005;51(8):1335‐1341. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Fig S1

Fig S2

Fig S3

Fig S4

Data Availability Statement

Not applicable.


Articles from Journal of Clinical Laboratory Analysis are provided here courtesy of Wiley

RESOURCES