Recommended supraventricular tachyarrhythmia discrimination settings

Parameter Recommended settings
Single-chamber devices
 MorphMatch ON
 Onset ON, 20%
 Stability ON, 40 ms
 Sustained VT timer OFF
DX, dual-chamber, and CRT devices
 SMART Detection ON (default settings or adapted to known VT)