Table II.

Analytical sensitivity of Truenat point-of-care (PoC) assays compared to reference rRT-PCR method values

Dilutions Truenat Beta CoV Truenat SARS-CoV-2 SARS-CoV 2 rRT-PCR



E gene RNase P RdRp-2 RNase P E gene RdRp-2 RNase P
Neat 15 20.33 13 20.5 19.29 24.49 26.45
D1 16.33 23.75 16.17 24.43 22.05 26.87 29.63
D2 20.6 28.11 18.5 27.86 25.78 30.28 33.01
D3 24 31.6 22.25 30.75 29.22 33.76 35.55 (-ve)
D4 27.2 33.29 25 33.6 32.4 36.87 (-ve) 38.6 (-ve)
D5 30.29 ND 28.8 ND 36.25 40.52 (-ve) 39.8 (-ve)
D6 33.33 ND 32 ND 38.3 ND ND