Skip to main content
. 2021 Jun 1;2021:3271395. doi: 10.1155/2021/3271395

Table 2.

Precision of the Sanger sequencing assay for TERTp mutations. Performance of interassay and intra-assay coefficients of variation.

Concentration n Total Within-day Between-day
Mean SD CV (%) Mean SD CV (%) Mean SD CV (%)
50% 30 1.14 0.12 10.72 1.21 0.02 1.84 1.20 0.12 10.00
15% 30 7.14 0.93 13.02 5.67 0.13 2.26 6.61 0.71 10.73

TERTp: the promoter region of the telomerase reverse transcriptase gene; SD: standard deviation; CV: coefficient of variation.