Table 3.
Detailed tests | Test results | Suggested tolerances | |
---|---|---|---|
Mechanical accuracy |
|
<1 mm | 1 mm |
|
<0.2 mm | 0.5 mm | |
|
<1.7 mm | 2 mm | |
Geometric accuracy |
|
<1 mm | 2 mm |
Patient setup accuracy |
|
<0.5 mm | 1 mm |
Image quality tests (regular and iCBCT reconstructions on all protocols) |
|
C: ≥5 lp/cm | C: not worse than 5 lp/cm |
Q: per protocol (0.25–0.68 cycles/mm) | Q: per protocol | ||
|
C: all HU inserts are visible | C: all HU inserts are visible | |
Q: per protocol (2–20 for regular CBCT) | Q: >2 | ||
|
C: ±30 HU | ±40 HU | |
Q: ±30 HU | |||
|
C: ±50 HU | ±50 HU | |
Q: ±50 HU | |||
|
C: <0.5 mm | 0.5 mm | |
Q: <0.5 mm | |||
|
<1 mm | 1 mm | |
Dosimetry |
|
<20 mGy within specs | Per protocol (1–40 mGy) |
|
<20 mGy within baseline | Per protocol (1–40 mGy) |
In the table, C represents the measurements with the Catphan phantom, and Q represents the measurement with the QUART phantom.