Skip to main content
. 2021 Jun 9;37:100954. doi: 10.1016/j.eclinm.2021.100954

Table 2.

Validation result of SARS-CoV-2 Rapid Antigen Test compared to RT-PCR test.

Asymptomatic Higher COVID-19 risk* Total sample
N 2288 255 2543
Cases (incidence) 15 (0·66) 34(13·33) 49 (1·93)
Sensitivity% (IC95%) 53·33 (26·59–78·73) 79·41 (62·10–91·30) 71·43 (56·74–83·42)
Specificity% (IC95%) 99·69 (99·37–99·88) 99·55 (97·50–99·99) 99·68 (99·37–99·86)
Positive predictive value% (CI95%) 53·33 (26·59–78·73) 96·43 (81·65–99·91) 81·40 (66·60–91·61)
Negative predictive value% (CI95%) 99·69 (99·37–99·88) 96·91 (93·74–98·75) 99·44 (99·06–99·69)

*Including participant with non-specific mild symptoms and close contacts.