Skip to main content
. 2021 Jun;42(6):1061–1068. doi: 10.3174/ajnr.A7082

Table 1:

Acquisition parameters for 3D-T2-STIR-ZOOMit and 2D T2-STIR

Parameter 3D-T2-STIR-ZOOMit 2D-T2-STIR
Mode 3D 2D
Acquisition plane Transversal Coronal
TR/TE (ms) 240/119 4880/101
Magnetization preparation Non-selective STIR Non-selective STIR
TI (ms) 220 220
FOV read/FOV phase 80 mm × 100% 180 mm × 100%
Base resolution/phase/section 128/100%/100% 320/100%/–
Acquired voxel size (X, Y, Z) (mm) 0.6, 0.6, 0.6 0.6, 0.6, 2.5
Interpolated voxel size (X, Y, Z) (mm) 0.3, 0.3, 0.6
Flip angle Variable 140°
Turbo factor 46 35
Averages 1.4 1
Acceleration factor (GRAPPA) 2 None
Scan duration (min/sec) 8/54 2/57

Note:—GRAPPA, generalized autocalibrating partially parallel acquisition; STIR, short tau inversion recovery.