Table 2. The results of H3-ECG detected the symptom-associated ECG data by automatic reading, compared to that of electrophysiologist-interpreted results.

Automatic reading Electrophysiologist-interpreted
Normal AF PAC PVC Total
Sensitivity and specificity were calculated for the simultaneous transmission (in bold). AF: sensitivity of 94% (47 of 50) and specificity of 98%; PACs: sensitivity of 93% (41 of 44) and specificity of 99%; PVCs: sensitivity of 98% (39 of 40) and specificity of 99%. AF: atrial fibrillation; PAC: premature atrial complex; PVC: prematureventricular complex.
Normal 97 1 0 0 98
AF 0 47 3 1 51
PAC 0 2 41 0 43
PVC 1 0 0 39 40
Total 98 50 44 40 232