Table 2.
FFR sample | Humidity | Heating cycles | Volunteer | Fit factor |
---|---|---|---|---|
01 | Ambient (<5% RH) | 1× | A | 107 (pass) |
02 | Ambient (<5% RH) | 1× | A | 195 (pass) |
03 | Ambient (<5% RH) | 1× | A | 165 (pass) |
04 | Ambient (<5% RH) | 1× | A | 149 (pass) |
05 | Ambient (<5% RH) | 10× | A | 200+ (pass)a |
06 | Ambient (<5% RH) | 10× | A | 200+ (pass) |
07 | Ambient (<5% RH) | 10× | A | 200+ (pass) |
08 | Ambient (<5% RH) | 10× | A | 200+ (pass) |
17 (control) | N/A | None | A | 181 (pass) |
09 | 90% RH | 1× | B | 200+ (pass) |
10 | 90% RH | 1× | B | 163 (pass) |
11 | 90% RH | 1× | B | 200+ (pass) |
12 | 90% RH | 1× | B | 200+ (pass) |
13 | 90% RH | 10× | B | 200+ (pass) |
14 | 90% RH | 10× | B | 200+ (pass) |
15 | 90% RH | 10× | B | 200+ (pass) |
16 | 90% RH | 10× | B | 161 (pass) |
18 (control) | N/A | None | B | 200+ (pass) |
aStrap broke upon doffing.