. 2021 Jun 12;141:104896. doi: 10.1016/j.jcv.2021.104896

Table 1.

Performance comparison of the ID NOW COVID-19 assay and a laboratory developed rRT-PCR assay to detect SARS-CoV-2 from paired or re-used ID NOW swabs.

Swab type rRT-PCR PPA (95% CI) NPA (95% CI)
Positive Negative
ID NOW (Paired ID NOW and MMM swabs)
Anterior nasal swab (n = 28)
Positive 6 0 100.0% 100.0%
Negative 0 22 (0.61–1.0) (0.85–1.0)
Midturbinate swab (n = 223)
Positive 18 1 94.7% 99.5%
Negative 1 203 (0.75–0.99) (0.97–0.99)
Nasopharyngeal swab (n = 92)
Positive 16 0 76.2% 100.0%
Negative 5 71 (0.54–0.89) (0.95–1.0)
Overall (n = 343)
Positive 40 1 86.9% 99.7%
Negative 6 296 (0.74–0.94) (0.98–0.99)
ID NOW (Re-used ID NOW swab in MMM)
Anterior nasal swab (n = 28)
Positive 6 0 100.0% 100.0%
Negative 0 22 (0.61–1.0) (0.85–1.0)
Midturbinate swab (n = 223)
Positive 18 1 90.0% 99.5%
Negative 2 202 (0.69–0.98) (0.97–0.99)
Nasopharyngeal swab (n = 92)
Positive 16 0 84.2% 100.0%
Negative 3 73 (0.68–0.94) (0.95–1.0)
Overall (n = 343)
Positive 40 1 88.8% 99.7%
Negative 5 297 (0.76–0.95) (0.98–0.99)