Table 6.
Rooms | Medical Care Act |
---|---|
Rooms | Medical radioisotope rooms |
Storage facilities | |
Disposal facilities | |
Radiotherapy rooms | |
Dose and concentration limits in controlled areas |
・Effective dose of external radiation: 1.3 mSv per 3 months ・Concentration of radioisotope (RI) in the air: The average concentration in 3 months is 1/10 of the concentration limit of the RI in the air ・Surface density of substances contaminated by RI: 1/10 of the surface density limit (alpha ray-emitting RI: 0.4 Bq/cm2, non-alpha ray-emitting RI: 4 Bq/cm2) |
Dose and concentration limits in regular thoroughfares in RI facilities |
・Effective dose outside the wall: 1 mSv per week ・RI concentration in air: the average weekly concentration is the RI concentration limit in air ・Surface density of substances contaminated by RI: Area density limit (alpha ray-emitting RI: 4 Bq/cm2, non-alpha ray-emitting RI: 40 Bq/cm2) |
Dose standards at the boundaries of hospitals (including residential areas inside the hospital) | Effective dose: 250 μSv per 3 months |
Exposure dose of in-patients | Effective dose should not exceed 1.3 mSv per 3 months |