Skip to main content
. 2021 Jun 12;13:38. doi: 10.1186/s13099-021-00435-3

Table 4.

Diagnostic performance of the 13C-UBT and 14C-UBT with optimal H. pylori-positive and -negative cutoff values, and the inclusion of an indeterminate range

Method Histology Sens (%)
95%CI
Spec (%)
95%CI
Acc (%)
95%CI
FPR (%)
95%CI
FNR (%)
95%CI
 + 
13C-UBT

 + 

(≥ 10.4%)

120 14

96

90.4–98.5

76.7

63.7–86.2

89.7

84.5–93.4

23.3

13.8–36.3

4

1.5–9.6

(≤ 2.9%)

5 46

IND

(3–10.3%)

18 14
14C-UBT

 + 

(≥ 238 DPM)

158 12

98.1

94.2–99.5

76.9

62.8–87.0

93

88.6–95.8

23.1

13.0–37.2

1.9

0.5–5.8

(≤ 86 DPM)

3 40

IND

(87–237 DPM)

31 23

UBT urea breath test, IND indeterminate, Sens sensitivity, Spec specificity, Acc accuracy, FPR false positive rate, FNR false negative rate, CI confidence interval