. 2021 May 5;22(6):229–240. doi: 10.1002/acm2.13221

Table 2.

Mean (±1SD) gamma pass rate difference between the original EPID dose image resolution of 0.255 mm (left columns) and 1mm resolution (right column) for lung SABR VMAT plans delivered with different collimator, MLCFS and MLCShift variations. Each plan consisted of two Arcs (Arc1 and Arc2). A lower gamma pass rate number indicates greater sensitivity to the introduced‐variation.

Variation type/magnitude

Gamma pass rate (%, mean ± SD)

EPID original image resolution of (0.255 mm)

Gamma pass rate (%, mean ± SD)

EPID image resolution of (1 mm)

3%/3 mm 2%/2 mm 2%/1 mm 1%/1 mm 3%/3 mm 2%/2 mm 2%/1 mm 1%/1 mm
Collimator (degree)
+1° Arc1 100.0 ± 0.0 100.0 ± 0.0 99.98 ± 0.0 99.66 ± 0.0 99.50 ± 0.0 97.37 ± 0.0 94.27 ± 0.0 88.51 ± 0.0
Arc2 100.0 ± 0.0 100.0 ± 0.0 100.0 ± 0.0 99.67 ± 0.0 99.64 ± 0.0 97.67 ± 0.0 88.88 ± 0.0 83.80 ± 0.0
+5° Arc1 100.0 ± 0.0 99.87 ± 0.1 93.27 ± 3.0 90.73 ± 2.4 99.64 ± 0.3 97.67 ± 2.0 88.88 ± 6.8 83.80 ± 8.6
Arc2 100.0 ± 0.0 99.90 ± 0.1 92.65 ± 2.5 90.85 ± 2.3 99.48 ± 0.1 95.67 ± 1.9 85.83 ± 6.2 81.47 ± 7.7
−2° Arc1 100.0 ± 0.0 100.0 ± 0.0 99.42 ± 0.0 98.88 ± 0.0 99.64 ± 0.0 96.32 ± 0.0 89.21 ± 0.0 83.87 ± 0.0
Arc2 100.0 ± 0.0 100.0 ± 0.0 99.34 ± 0.0 98.65 ± 0.0 99.64 ± 0.0 96.47 ± 0.0 89.16 ± 0.0 83.61 ± 0.0
−5° Arc1 100.0 ± 0.0 98.82 ± 1.3 89.85 ± 1.3 87.17 ± 2.3 99.31 ± 0.2 92.43 ± 1.1 79.64 ± 1.6 73.98 ± 3.5
Arc2 100.0 ± 0.0 99.08 ± 1.3 89.75 ± 3.7 87.40 ± 4.6 99.28 ± 0.4 92.91 ± 1.6 79.63 ± 4.2 73.95 ± 6.3
MLCFS (mm)
+1 Arc1 100.0 ± 0.0 100.0 ± 0.0 96.77 ± 0.9 92.51 ± 1.8 99.47 ± 0.3 96.90 ± 0.8 77.32 ± 3.2 73.85 ± 4.3
Arc2 100.0 ± 0.0 100.0 ± 0.0 98.13 ± 0.9 93.83 ± 1.7 99.68 ± 0.2 97.43 ± 0.9 77.62 ± 3.6 74.43 ± 4.4
+2 Arc1 100.0 ± 0.0 97.95 ± 0.0 79.02 ± 0.0 76.18 ± 0.0 99.32 ± 0.0 85.82 ± 0.0 74.12 ± 0.0 70.70 ± 0.0
Arc2 99.96 ± 0.0 98.81 ± 0.0 78.41 ± 0.0 74.30 ± 0.0 99.99 ± 0.0 86.85 ± 0.0 73.74 ± 0.0 68.76 ± 0.0
+5 Arc1 81.36 ± 0.0 68.76 ± 0.0 64.05 ± 0.0 60.61 ± 0.0 78.08 ± 0.0 67.12 ± 0.0 63.70 ± 0.0 59.87 ± 0.0
Arc2 79.12 ± 0.0 67.73 ± 0.0 64.52 ± 0.0 61.33 ± 0.0 75.77 ± 0.0 67.15 ± 0.0 63.84 ± 0.0 59.36 ± 0.0
−1 Arc1 100.0 ± 0.0 100.0 ± 0.1 96.39 ± 1.2 92.47 ± 1.9 99.33 ± 0.5 96.55 ± 1.2 77.71 ± 3.6 73.63 ± 3.6
Arc2 100.0 ± 0.0 99.92 ± 0.2 96.14 ± 3.0 92.84 ± 2.9 99.62 ± 0.2 97.09 ± 1.0 77.92 ± 4.0 74.32 ± 4.1
−2 Arc1 100.0 ± 0.0 96.36 ± 0.0 79.75 ± 0.0 75.10 ± 0.0 99.53 ± 0.0 84.24 ± 0.0 75.71 ± 0.0 69.64 ± 0.0
Arc2 100.0 ± 0.0 96.62 ± 0.0 78.67 ± 0.0 73.98 ± 0.0 95.56 ± 0.0 84.56 ± 0.0 72.50 ± 0.0 66.73 ± 0.0
MLC shift (mm)
+1 Arc1 100.0 ± 0.0 100.0 ± 0.0 95.07 ± 4.3 91.13 ± 5.4 99.46 ± 0.4 96.64 ± 1.1 77.97 ± 3.6 74.62 ± 4.0
Arc2 100.0 ± 0.0 100.0 ± 0.0 96.94 ± 0.7 91.83 ± 2.5 99.75 ± 0.2 97.54 ± 0.9 78.54 ± 3.9 75.52 ± 4.3
+2 Arc1 100.0 ± 0.0 98.39 ± 0.0 73.63 ± 0.0 69.23 ± 0.0 99.61 ± 0.0 82.64 ± 0.0 69.06 ± 0.0 64.21 ± 0.0
Arc2 100.0 ± 0.0 99.21 ± 0.0 75.86 ± 0.0 70.08 ± 0.0 99.43 ± 0.0 85.14 ± 0.0 69.32 ± 0.0 63.25 ± 0.0
+5 Arc1 78.64 ± 0.0 69.26 ± 0.0 64.90 ± 0.0 62.71 ± 0.0 75.92 ± 0.0 67.99 ± 0.0 64.79 ± 0.0 62.09 ± 0.0
Arc2 77.09 ± 0.0 67.79 ± 0.0 64.21 ± 0.0 60.75 ± 0.0 74.64 ± 0.0 67.18 ± 0.0 63.89 ± 0.0 60.09 ± 0.0
−1 Arc1 100.0 ± 0.0 100.0 ± 0.0 97.50 ± 3.6 94.70 ± 5.0 99.43 ± 0.0 96.33 ± 0.6 78.51 ± 3.2 74.83 ± 4.4
Arc2 100.0 ± 0.0 100.0 ± 0.0 97.36 ± 1.8 92.19 ± 1.7 99.66 ± 0.2 97.34 ± 1.0 78.72 ± 4.4 75.67 ± 4.9
−2 Arc1 100.0 ± 0.0 94.70 ± 3.5 74.87 ± 2.0 70.72 ± 2.8 99.69 ± 0.0 82.13 ± 1.0 71.50 ± 2.8 67.50 ± 3.6
Arc2 100.0 ± 0.0 94.20 ± 3.2 75.90 ± 0.9 72.26 ± 1.6 99.58 ± 0.4 82.41 ± 1.0 72.37 ± 1.8 67.46 ± 3.5