. 2021 May 5;22(6):183–190. doi: 10.1002/acm2.13263

Table 1.

Measurements of inner diameter, wall thickness, and outer diameter for the five simulated airways tubes, from A (largest) to E (smallest) from micro‐CT and clinical CT scanning (scanner A). Average variation denotes measured difference from nominal values. Tube E was not well visualized with clinical CT and hence not measured.

Tube Measurement Designed Nominal value (mm) Micro‐CT average ± std. dev. (mm) Micro‐CT average variation from nominal (mm) Clinical CT average ± std. dev. (mm) Clinical CT average variation from nominal (mm)
A Inner Diameter 5 4.98 ± 0.021 −0.016 4.60 ± 0.013 −0.40
Wall Thickness 2.3 2.33 ± 0.027 0.028 2.50 ± 0.021 0.20
Outer Diameter 9.6 9.64 ± 0.034 0.039 9.60 ± 0.035 0.004
B Inner Diameter 4 3.97 ± 0.021 −0.031 3.45 ± 0.032 −0.55
Wall Thickness 1.8 1.83 ± 0.026 0.028 2.07 ± 0.042 0.27
Outer Diameter 7.6 7.63 ± 0.035 0.026 7.60 ± 0.066 −0.002
C Inner Diameter 3 2.95 ± 0.025 −0.052 2.37 ± 0.042 −0.63
Wall Thickness 1.3 1.35 ± 0.024 0.046 1.59 ± 0.018 0.29
Outer Diameter 5.6 5.64 ± 0.038 0.039 5.54 ± 0.049 −0.058
D Inner Diameter 2 1.91 ± 0.022 −0.086 1.31 ± 0.033 −0.69
Wall Thickness 0.8 0.85 ± 0.022 0.047 1.01 ± 0.028 0.21
Outer Diameter 3.6 3.61 ± 0.041 0.008 3.32 ± 0.042 −0.28
E Inner Diameter 1 0.92 ± 0.023 −0.080 NA NA
Wall Thickness 0.4 0.44 ± 0.022 0.035 NA NA
Outer Diameter 1.8 1.79 ± 0.029 −0.010 NA NA