Table 1.
Levels of the parameter for each beam model defined for the design of experiment.
| Level | σX (mm) | σY (mm) | DLG (mm) | T (%) | |
|---|---|---|---|---|---|
| TrueBeam Tx ‐ Millennium MLC ‐ X6 | 1 | 0.7 | 0 | 0.75 | 1 |
| 2 | 1.3 | 1.1 | 2.3 | 1.8 | |
| 3 | 1.8 | 1.4 | 3 | 2 | |
| TrueBeam STx ‐ 120 HD MLC ‐ X6 | 1 | 0 | 0 | 0.2 | 0.8 |
| 2 | 0.6 | 0.6 | 1.5 | 1.2 | |
| 3 | 1.2 | 1.2 | 2.5 | 1.8 | |
| TrueBeam STx ‐ 120 HD MLC ‐ X6 FFF | 1 | 0 | 0 | 0 | 0.8 |
| 2 | 0.6 | 0.6 | 0.8 | 1.2 | |
| 3 | 1.2 | 1.2 | 1.6 | 1.8 |
σX/σY, effective target spot size in the X/Y direction; DLG, dosimetric leaf gap; T, transmission.