Table 3.
Significant parameters on the field output factors (parameters are ordered with decreasing significance).
| 0.5 cm | 1 cm | 1.5 cm | 2 cm | 3 cm | 4 cm |
|---|---|---|---|---|---|
| σY | DLG | DLG | DLG | DLG | DLG |
| DLG | σY | σY | T | T | T |
| σX | σX | σX | σY | ||
| DLG‐σX interaction | T | T | σX a | ||
| T | DLG‐σX interaction | DLG‐σX interaction b | |||
| DLG‐σY interaction a |
T, transmission.
Significant for two of the three beam models.
Significant for one of the beam models.