Number of diagnostic-quality images, handheld Butterfly vs cart-based GE ultrasound device.

Component of UHP exam Butterfly N = 29
N (%, 95% CI)
GE N = 29
N (%, 95% CI)
Absolute difference 95% CI P-value*
PLAX 27 (93, 77 – 99) 28 (97, 82 – 100) −3 (−15 – 8) 1.00
4C View 26 (90, 73 – 98) 25 (86, 68 – 96) 3 (−13 – 20) 1.00
IVC 23 (79, 60 – 92) 24 (82, 64 – 94) −3 (−23 – 17) 1.00
RUQ 29 (100, 88 – 100) 29 (100, 88 – 100) 0 (0 – 0)
LUQ 29 (100, 88 – 100) 29 (100, 88 – 100) 0 (0 – 0)
Bladder 29 (100, 88 – 100) 29 (100, 88 – 100) 0 (0 – 0)
Aorta 25 (86, 68 – 96) 23 (79, 68 – 96) 7 (−12 – 26) 0.73
RL 29 (100, 88 – 100) 29 (100, 88 – 100) 0 (0 – 0)
LL 29 (100, 88 – 100) 29 (100, 88 – 100) 0 (0 – 0)

UHP, ultrasound hypotension protocols; CI, confidence interval; PLAX, parasternal long axis; 4C, four chamber; IVC, inferior vena cava; RUQ, right upper quadrant; LUQ, left upper quadrant; RL, right leg; LL, left leg.