Skip to main content
. 2021 Apr 8;22(3):775–781. doi: 10.5811/westjem.2020.12.47862

Table 3.

Comparison of task-switching time, handheld Butterfly vs GE cart-based ultrasound device.

Task switching Butterfly GE Difference 95% CI P-value
Total time 1
 Mean, (SD) 5.55 (3.51) 9.57 (3.73) −4.02 (−5.93 – −2.12) <0.0001
 Median, (IQR) 4.28 (3.63 – 5.62) 8.52 (7.17 – 9.57) −4.23 (−5.48 – −2.98) <0.0001
Total time 2
 Mean, (SD) 5.32 (3.54) 7.13 (3.51) −1.81 (−3.67 – 0.04) 0.0556
 Median, (IQR) 4.08 (3.43 – 5.35) 5.8 (4.73 – 7.67) −1.72 (−2.78 – −0.65) <0.0001
Switching 1
 Mean, (SD) 0.13 (0.06) 0.30 (0.08) −0.17 (−0.21 – −0.13) <0.0001
 Median, (IQR) 0.13 (0.08 – 0.17) 0.30 (0.23 – 0.35) −0.17 −0.22 – −0.12 <0.0001
Switching 2
 Mean, (SD) 0.09 (0.08) 0.08 (0.10) 0.01 (−0.04 – 0.06) 0.6893
 Median, (IQR) 0.08 (0.00 – 0.13) 0.00 (0.0 – 0.2) 0.08 (−0.02 – 0.18) 0.0825
Switching 3
 Mean, (SD) 0.12 (0.05) 0.33 (0.23) −0.20 (−0.30 – −0.11) <0.0001
 Median, (IQR) 0.10 (0.10 – 0.13) 0.28 (0.23 – 0.38) −0.18 (−0.23 – −0.14) <0.0001
Total switching
 Mean, (SD) 0.34 (0.14) 0.71 (0.31) −0.37 (−0.37 – −0.50) <0.0001
 Median, (IQR) 0.33 (0.23 – 0.42) 0.67 (0.48 – 0.82) −0.33 (−0.45 – −0.21) <0.0001

Total time 1 (minutes): includes transport and plug-in.

Total time 2 (minutes): excludes transport and plug-in.

Switching 1 (minutes): cardiac to abdomen preset.

Switching 2 (minutes): abdomen to abdomen deep preset.

Switching 3 (minutes): abdomen deep to lung preset.