TABLE 1.

MR sequence parameters used for (A) 1.5T and (B) 3T imaging protocols.

(A) 1.5T (Signa HDxt)
Sequence GE-EPI BOLD 3D IR-SPGR T1-w T2-w Propeller FLAIR GRE
TR (ms) 3,000 9.6 7,000 8,000 900
TE (ms) 45 4.0 90 100 15
TI (ms) 500 2,000
Flip angle (°) 90 8 20
Field of view (cm) 25.6 × 25.6 25.6 × 25.6 24 × 24 24 × 24 24 × 24
Acquisition matrix 64 × 64 192 × 192 320 × 256 384 × 256
Slice thickness (mm) 4 1.3 4 4 4
Number of slices 36 160 36 36 36
Other 1.5 signal averages, matrix size 384

(B) 3T (Siemens Verio)
Sequence GE-EPI BOLD 3D MPRAGE T1-w T2-w BLADE FLAIR BLADE GRE

TR (ms) 3,000 2,300 11,400 9,100 600
TE (ms) 30 2.98 120 125 10
TI (ms) 1,100 2,512
Flip angle (°) 90 9 130 15
Field of view (cm) 19.2 × 19.2 25.6 × 25.6 24 × 24 24 × 24 24 × 24
Acquisition matrix 64 × 64 256 × 256 384 × 384 256 × 256 320 × 256
Slice thickness (mm) 3 1 3 3 3
Number of slices 48 256 48 48 48
Other

GE-EPI, gradient-echo echo-planar imaging; IR-SPGR, inversion recovery spoiled gradient echo; MPRAGE, magnetization prepared-rapid gradient echo; FLAIR, fluid-attenuated inversion recovery; GRE, gradient recalled echo.