Table 2.

Diagnostic performance of TE and SWE-10, SWE-5, or SWE-3 for the diagnosis of significant (F2-F4) and advanced fibrosis (F3-F4)

AUROC 95% CI P-value
Significant fibrosis
 TE 0.719 0.572-0.867 0.016
 SWE-10 0.787 0.646-0.927 0.002
 SWE-5 0.809 0.676-0.942 0.001
 SWE-3 0.714 0.560-0.869 0.021
 P-value 0.301
Advanced fibrosis
 TE 0.799 0.646-0.952 <0.001
 SWE-10 0.797 0.659-0.935 <0.001
 SWE-5 0.809 0.684-0.933 <0.001
 SWE-3 0.736 0.587-0.885 0.003
 P-value 0.496

TE, transient elastography; SWE-10, p-SWE with 10 measurements; SWE-5, p-SWE with five measurements; SWE-3, p-SWE with three measurements; p-SWE, point shear-wave elastography; AUROC, area under the receiving operator characteristic curve; CI, confidence interval.