. 2021 Jun 18;59(7):e02942-20. doi: 10.1128/JCM.02942-20

TABLE 5.

Diagnostic accuracy of the iNtRON PCR kit, using a ≥4-fold rise in IFA antibody titers or an IFA IgM/IgG cutoff value of ≥1:1,024 as a gold standarda

Rising IFA antibody titer or IFA IgM/IgG cutoff value of ≥1,024
Rising IFA antibody titer alone
Total
Before antibiotic
After antibiotic
Total
Before antibiotic
After antibiotic
Parameter Case Control Case Control Case Control Case Control Case Control Case Control
No. of iNtRON kit results:
 Positive 70 0 49 0 17 0 60 0 43 0 13 0
 Negative 18 51 9 17 9 1 9 44 6 16 3 1
 Total 88 51 58 17 26 1 69 44 48 16 16 1
Sensitivity, % (95% CI) 79.5 (69.3–87.1) 84.5 (72.1–92.2) 65.4 (44.4–82.1) 87.0 (76.2–93.5) 87.8 (74.5–94.9) 81.3 (53.7–95.0)
Specificity, % (95% CI) 100 (91.3–100) 100 (77.1–100) 100 (5.5–100) 100 (90.0–100) 100 (75.9–100) 100 (5.5–100)
PPV, % (95% CI) 100 (93.5–100) 100 (90.9–100) 100 (77.1–100) 100 (92.5–100) 100 (89.8–100) 100 (71.7–100)
NPV, % (95% CI) 73.9 (61.7–83.4) 65.4 (44.4–82.1) 10.0 (0.5–45.9) 83.0 (69.7–91.5) 72.7 (49.6–88.4) 25 (1.3–78.1)
a

IFA, indirect immunofluorescence assay; 95% CI, 95% confidence interval; PPV, positive predictive value; NPV, negative predictive value.