Results of Cyclic and Load-to-Failure Testing (Mechanical Study Part) of 9 Different High-Strength Sutures and Tapes

Product N Stiffness, N/mm
Force at 3 mm, N
Displacement at 200 N, mm
Ultimate Load to Failure, N
Displacement at Failure, mm
Mean SD Mean SD Mean SD Mean SD Mean SD
FiberTape 6 50 5.2 376 19.1 2.0 0.10 806 36.2 8.8 3.78
Hi-Fi Tape 6 56 3.2 205 23.2 3.2 0.74 592 52.4 25.6 5.79
Permatape 6 58 5.3 219 11.0 2.8 0.12 551 38.1 7.4 1.66
SutureTape 6 37 2.3 130 14.6 4.2 0.34 327 24.7 5.9 0.35
UltraTape 6 54 4.9 197 35.0 5.7 0.39 402 37.6 5.9 1.11
XBraid TT 6 23 3.2 80 11.8 6.7 0.75 378 69.9 10.9 2.58
Dynacord 6 9 0.5 19 1.8 10.0 0.58 314 21.6 12.4 1.47
FiberWire 6 42 4.4 151 9.0 4.1 0.41 258 15.3 6.0 0.98
Ultrabraid 6 24 0.9 82 10.6 5.7 0.39 294 28.3 7.6 0.66
Total 54 29 2.3 162 100.5 4.6 2.42 436 174.5 10.1 6.43