Table 3.

Comparison of Results of Cyclic and Load-to-Failure Testing of High-Strength Sutures and Tapes in Human Rotator Cuff Tendons (Cadaveric Study Part)

Parameter Tapes (n = 48) Sutures (n = 24) P Value
Stiffness, N/mm 20.4 ± 7.6 16.9 ± 5.1 .029
Cyclic displacement, mm 0.8 ± 0.65 0.6 ± 0.33 .586
Force at 3 mm of displacement, N 55.3 ± 25.4 43.6 ± 16.8 .029
Displacement at 50 N, mm 2.6 ± 1.13 3.1 ± 1.01 .113
Displacement at 100 N, mm 6.2 ± 2.56 7.1 ± 2.48 .257
Ultimate load to failure, N 153 ± 61.5 153 ± 67.5 .996
Knot height, mm 7.6 ± 1.55 6.4 ± 1.63 .007
Knot width, mm 2.3 ± 0.82 1.7 ± 0.31 .001

NOTE. Data are presented as mean ± standard deviation.