Table 4.

Results of Cyclic and Load-to-Failure Testing of 9 Different High-Strength Sutures and Tapes in Human Rotator Cuff Tendons (Cadaveric Study Part)

Product N Displacement, mm
Ultimate Load to Failure, N
Stiffness, N/mm
Knot Height, mm
Knot Width, mm
Force at 3 mm of Displacement, N
Displacement at 100 N, mm
Displacement at 50 N, mm
Mean SD Mean SD Mean SD Mean SD Mean SD Mean SD Mean SD Mean SD
FiberTape 8 0.4 0.38 178 50.5 26 7.6 9.5 1.28 3.8 0.69 85 29.4 4.5 3.18 1.9 0.65
Hi-Fi Tape 8 0.8 0.36 134 72.6 20 7.5 8.2 0.79 2.6 0.37 42 17.9 6.1 2.02 2.3 0.85
Permatape 8 0.8 0.59 166 72.1 21 10.6 7.4 1.18 2.0 0.35 51 23.9 5.6 1.47 2.4 0.81
SutureTape 8 0.9 1.18 174 48.1 20 6.9 5.6 0.67 1.6 0.28 56 19.7 6.5 1.66 2.6 1.39
UltraTape 8 0.9 0.24 104 37.2 21 5.6 7.4 0.99 2.2 0.46 54 21.3 6.8 2.72 3.5 0.98
XBraid TT 8 1.0 0.57 156 72.7 14 3.2 7.8 1.36 2.0 0.23 38 10.2 8.2 2.81 3.7 1.07
Dynacord 8 0.9 0.27 130 67.1 15 6.5 6.3 0.97 1.9 0.31 37 15.4 8.8 3.35 3.5 1.05
FiberWire 8 0.7 0.16 173 65.3 20 3.4 6.0 0.68 1.6 0.39 51 20.1 5.6 0.73 2.3 0.38
Ultrabraid 8 0.9 0.48 158 72.1 16 2.9 6.8 2.58 1.6 0.19 43 13.4 7.1 2.07 3.5 0.98
Total 72 0.8 0.56 153 63.1 19 7.0 7.2 1.67 2.1 0.75 51 23.5 6.5 2.54 2.8 1.11

SD, standard deviation.