Skip to main content
. 2021 Jun 8;33(2):mzab091. doi: 10.1093/intqhc/mzab091

Table 1.

Number and size of glass particles by ampoule size and breaking method (n = 672)

Number of glass particles
Breaking method Ampoule size (ml) <20G
<60 µm
19G
61–68 µm
18G
69–83 µm
84–120 µm >120 µm Total Mean (SD)
Method 1 10 246 5 2 13 22 288 5.14 (4.97)
2 142 1 3 7 11 164 2.93 (4.24)
388 6 5 20 33 452 4.04 (4.73)
Method 2 10 401 6 8 21 46 482 8.61 (6.37)
2 173 0 3 2 11 189 3.38 (4.06)
574 6 11 23 57 671 5.99 (5.93)
Method 3 10 189 1 1 6 14 211 3.77 (4.06)
2 47 0 0 2 9 59 1.05 (2.08)
236 1 2 8 23 270 2.41 (3.99)
Method 4 10 219 0 6 18 19 262 4.68 (5.76)
2 64 0 2 7 1 74 1.32 (1.97)
283 0 8 25 20 336 3.00 (4.61)
Method 5 10 235 2 5 12 28 282 5.04 (4.26)
2 141 2 2 6 6 157 2.80 (5.68)
376 4 7 18 34 439 3.92 (5.12)
Method 6 10 343 0 4 12 23 382 6.82 (6.86)
2 175 0 0 3 15 194 3.46 (6.27)
518 0 5 15 38 576 5.14 (6.75)
Total 10 1633 14 26 82 152 1907 5.68 (5.04)
2 742 3 12 27 53 837 2.99 (4.44)