Table 1.

Clinical performance of AccuPower SARS-CoV-2 Multiplex RT-PCR kit and Allplex 2019-nCoV Assay using the CDC protocol as a gold standard (% values: sensitivity).

RT-PCR kit Positive samples (including “presumptive positive: samples) False-negative samples Total SARS-CoV-2 positive samples
Allplex 2019-nCoV Assay 55 (96.5%) 2 57
AccuPower SARS-CoV-2 Multiplex RT-PCR kit 37 (75.5%) 12 49

Only SARS-CoV-2 positive samples included on the study are detailed.