Clinical performance of DBS vs. plasma for PCR-confirmed COVID-19 detection.

Comparisons Threshold (U/mL) True
Negatives
True
Positives
False
Negatives
False
Positives
Sensitivity Specificity
Plasma vs. PCR-confirmed COVID-19 0.8 11 46 6 0 0.88 (95% CI: 0.77, 0.96) 1 (95% CI: 0.72, 1)
DBS vs. PCR-confirmed COVID-19 0.8 11 45 7 0 0.87 (95% CI: 0.74, 0.94) 1 (95% CI: 0.72, 1)