Skip to main content
. 2021 Jun 24;11:13233. doi: 10.1038/s41598-021-92338-1

Table 2.

A Inter-day intermediate precision of PMF-EIC method with mAb X containing ~ 0.1% of E262K substitution. The % substitution is calculated using Eq. (1). B Accuracy of PMF-EIC method at ~ 400 pmol and ~ 8 pmol column load.

Intermediate inter-day precision % E262K
A
Day 1 0.08
Day 2 0.12
Day 3 0.14
Day 4 0.10
Day 5 0.12
Day 6 0.09
Average 0.11
% CV 18.8
Column load AUC of EK peptide AUC of native peptide %E262K Average %E262K % CV
B
  ~ 400 pmol 130,195,529 331,971,597 28.2 28.8 5.7
128,506,560 296,820,826 30.2
131,613,201 339,193,814 28.0
  ~ 8 pmol 3,463,371 2,633,558 56.8 47.5 44.3
1,958,042 3,970,226 33.0
1,918,275 1,717,359 52.8