. 2021 May 12;7(6):963–972. doi: 10.1021/acscentsci.1c00186

Figure 4.

Figure 4

Loop-mediated isothermal amplification (LAMP) lower limit of detection study. (a) 12% acrylamide gel of digested HPV 18 LAMP amplicons; stochastic amplification occurs below 104 cp/μL of input HPV 18 plasmid DNA. (b) Only samples that contained positive HPV 18 LAMP amplicons, as determined by gel electrophoresis, resulted in a signal decrease in methylene blue on the gold leaf electrodes below a given threshold. (c) Probit analysis of the electrochemical data indicates that the lower limit of detection for the electrochemical detection of LAMP amplicons is 1.2 × 104 total copies of HPV 18 DNA.