Table 3.
Clinical parameters of the studied subjects
| Parameter | ULD CT-Pure IR | PRSS | p-value |
|---|---|---|---|
| Demonstration of exact shunt entry point to peritoneum/pleura/atrium | 22/22 (100%) | Not testable | - |
| Clear demonstration of shunt tip | 22/22 (100%) | 21/22 (95.5%) | p > 0.009 |
| Precise identification of anatomic location of shunt tip | 22/22 (100%) | Not testable | - |
| Visualisation of entire imaged shunt course | 22/22 (100%) | 8/22 (36.4%) | p < 0.001 |
| Demonstration of shunt complications | Not testable * | Not testable * | - |
PRSS and ULD CT-Pure IR images were compared using a binary scale (1, yes; 2, no) for the included parameters. Contingency squares were generated for each variable, and Fisher’s exact test was performed. A p-value < 0.05 was considered statistically significant
*There were no shunt-related complications in this study