Table 2.
Subgroup analysis in function of KardiaMobile numerical precision compared to conventional electrocardiogram
| Characteristics | N | Numerical precision† |
P | |
|---|---|---|---|---|
| Cases | Percentage | |||
| Number | 128 | 119 | 93.0% | |
| Sex | ||||
| Male | 61 | 56 | 91.8% | |
| Female | 67 | 63 | 94.0% | .623 |
| SARS-CoV-2 infection | ||||
| Probable and confirmed | 40 | 34 | 85.0% | |
| Ruled out | 88 | 85 | 96.6% | .017‡ |
| Heart failure history | ||||
| Yes | 53 | 49 | 92.5% | |
| No | 75 | 70 | 93.3% | .848 |
| Blood hypertension history | ||||
| Yes | 78 | 74 | 94.9% | |
| No | 50 | 45 | 90.0% | .293 |
| Potentially QT-prolonging drugs | ||||
| Yes | 65 | 60 | 92.3% | |
| No | 63 | 59 | 93.7% | .766 |
Data are presented as number and percentage.
P value obtained through χ2 test.
Numerical precision was defined as the proportion of KardiaMobile (AliveCor, San Francisco, CA) QT interval measurements with less than 10 ms difference from conventional electrocardiogram measurements.
P < .05.