Table 2.

Troubleshooting guide for SCFA detection with GC-MS.

Problem Possible Cause(s) Solution
Vacuum pressure is consistently higher than that recommended for flow rate Air leaks or other vacuum problems Consult operation manual
Spurious chromatogram peaks and poor repeatability Backflash in response to flooding the injection line Inject less sample volume
Poor recovery of analyte from sample Ensure that freeze/thaw cycles have been kept to ≤ 2 and that samples are stored at −80 °C for long-term storage
Shifting retention, baseline disturbances Column bleed, column contamination, column aging Replace with new column and condition before use