Skip to main content
. 2021 Jun 28;11:13378. doi: 10.1038/s41598-021-92677-z

Table 1.

Limit of detection determined by dilution series for Saliva-Dry LAMP using commercially-available instruments and the Biobox manufactured in this study.

Sample concentration (copies/μL) Positive reactions
Commercially-available instruments Biobox
1.0 4/4 4/4
0.5 4/4 3/4
0.25 0/4 2/4

A single experiment performed in quadruplicate is shown at each concentration using a contrived saliva sample containing SARS-CoV-2.