Table 3.

Urea clearance achieved from different dialyzers at different flow rates[29,47,51]

Manufacturer Brand of Dialyzer Membrane Material Membrane Surface Area (m2) K (urea) in ml/min when QB is 200 ml/min, and QD is 500 ml/min K (urea) in ml/min when QB is 400 ml/min, and QD is 500 ml/min
Baxter Revaclear 300 PAES/PVP 1.4 196 323
Baxter Revaclear 400 PAES/PVP 1.8 198 338
Fresenius Hemoflow F6 Fresenius Polysulfone® polymer 1.3 171 252
Fresenius Hemoflow F8 Fresenius Polysulfone® polymer 1.8 177 280
Fresenius Optiflux F180A; F180B Fresenius Polysulfone® polymer 1.8 194 314
Fresenius Optiflux F200A, F200B Fresenius Polysulfone® polymer 2.0 194 320

QB is blood flow rate, QD is dialysate flow rate, K is clearance, PAES is Polyarylethersulfone, PVP is Polyvinylpyrrolidone