Skip to main content
. 2021 Jun 24;29:152–162. doi: 10.1016/j.jot.2021.05.009

Table 2.

The deviation errors of computer navigation-assisted bone resection in long bone sarcoma.

Studies Number of patients (osteotomies) Mean age Bone reconstruction Mean maximum deviation errors
Aponte-Tinao et al. (2013) [11] 5 patients (18 CN) 56 (42–71) Allograft 2.43 ​± ​1.8 ​mm
Bosma et al. (2018) [20]: Cadaveric study 8 procedures (14 CN) NA NA 3.6 ​± ​2.1 ​mm
8 procedures (14 NAVIG) NA NA 2.0 ​± ​1.0 ​mm
Zhang et al. (2020) [26] 10 patients (16 CN) 26.6 (12–52) Not mentioned 2.52 ​± ​1.38 ​mm
Current study 3 patients (9 NAVIG) 13.7 (8–17) Custom Patient-specific implants 1.64 ​± ​0.35 ​mm

CN: computer navigation assistance; NAVIG: Computer Navigation and patient-specific cutting Guide; N/A: not applicable