Skip to main content
. 2021 Jul 18;14(7):1013–1017. doi: 10.18240/ijo.2021.07.08

Table 1. Comparison of biometric measurements of adolescents with ametropia by two biometric instruments.

Instrument No. of eyes AL (mm) Km (D) ACD (mm) WTW (mm)
IOL-Master700 110 24.805±0.969 42.642±1.365 3.777±0.224 12.138±0.415
IOL-Master500 110 24.777±0.959 42.774±1.295 3.676±0.219 12.075±0.383
difference 110 0.028±0.110 -0.102±0.674 0.101±0.085 0.064±0.171
t 2.664 -1.583 12.505 3.911
P 0.009 0.116 0.000 0.000

AL: Axial length; Km: Mean corneal anterior surface keratometry; ACD: Anterior chamber depth; WTW: Horizontal corneal diameter.

mean±SD