Table 5: NTA results of human plasma derived EVs demonstrate the effect of altering laser power and gain on size and particle concentration measurements of human EVs (1:1000 dilution).

D10 is the point in the size distribution where 10% of the sample is contained, D50 is the point where 50% of the sample is contained (median), and D90 is the point where 90% of the sample is contained. SD: Standard deviation; CV: Coefficient of variation.

Gain [dB] Blue Laser [mW] Green Laser [mW] Red Laser [mW] Total Concentration, particles/mL (raw) Average size [nm]* Standard deviation of size / CV Modal size D10 / D50 / D90
18 70 12 8 5.90E+07 133 89 / 0.66 67 66.57 / 131.08 / 322.70
24 70 12 8 1.00E+08 118 73 / 0.61 64 60.21 / 117.93 / 257.08
30 210 12 8 1.70E+08 105 57 / 0.54 80 59.83 / 104.74 / 206.07
30 70 12 8 1.10E+08 122 75 / 0.61 74 73.17 / 123.09 / 278.70
30 210 0 0 1.00E+08 116 70 / 0.6 82 71.33 / 115.63 / 257.65
30 70 0 0 7.00E+07 126 79 / 0.63 82 74.81 / 125.16 / 284.23
30 0 12 0 2.80E+07 169 106 / 0.63 100 98.28 / 163.39 / 392.50
30 0 24 0 4.40E+07 148 95 / 0.64 88 84.24 / 147.69 / 354.15
30 0 0 8 1.50E+07 175 129 / 0.74 62 8.81 / 15.32 / 108.93
30 0 0 16 9.20E+06 246 147 / 0.6 13 8.55 / 14.93 / 136.75
*

Integrated within 50–1000 nm range