Abstract
The aim of this work was to improve the SNR efficiency of zero echo time (ZTE) MRI pulse sequences for faster imaging of short‐T 2 components at large dead‐time gaps. ZTE MRI with hybrid filling (HYFI) is a strategy for retrieving inner k‐space data missed during the dead‐time gaps arising from radio‐frequency excitation and switching in ZTE imaging. It performs hybrid filling of the inner k‐space with a small single‐point‐imaging core surrounded by a stack of shells acquired on radial readouts in an onion‐like fashion. The exposition of this concept is followed by translation into guidelines for parameter choice and implementation details. The imaging properties and performance of HYFI are studied in simulations as well as phantom, in vitro and in vivo imaging, with an emphasis on comparison with the pointwise encoding time reduction with radial acquisition (PETRA) technique. Simulations predict higher SNR efficiency for HYFI compared with PETRA at preserved image quality, with the advantage increasing with the size of the k‐space gap. These results are confirmed by imaging experiments with gap sizes of 25 to 50 Nyquist dwells, in which scan times for similar image quality could be reduced by 25% to 60%. The HYFI technique provides both high SNR efficiency and image quality, thus outperforming previously known ZTE‐based pulse sequences, in particular for large k‐space gaps. Promising applications include direct imaging of ultrashort‐T2 components, such as the myelin bilayer or collagen, T 2 mapping of ultrafast relaxing signals, and ZTE imaging with reduced chemical shift artifacts.
Keywords: bone, gap filling, PETRA, short T2, SNR efficiency, SPI, WASPI, ZTE
HYFI is a strategy for retrieving inner k‐space data missed during the dead‐time gap arising from radio‐frequency excitation and switching in ZTE imaging. It performs hybrid filling of the inner k‐space with a small single‐point‐imaging core surrounded by a stack of shells and provides both high SNR efficiency and image quality, thus outperforming previously known ZTE‐based pulse sequences, particularly at large dead‐time gaps. Promising applications include direct imaging of ultrashort components, such as the myelin bilayer or collagen.
Abbreviations used
- FID
free induction decay
- FOV
field of view
- HYFI
ZTE MRI with hybrid filling
- MTF
modulation transfer function
- PETRA
pointwise encoding time reduction with radial acquisition
- PSF
point spread function
- RHE
ramped hybrid encoding
- SNR
signal‐to‐noise ratio
- SPI
single‐point imaging
- SWIFT
sweep imaging with Fourier transformation
- ZTE
zero echo time
1. INTRODUCTION
Direct MRI of tissues with very short transverse relaxation times T 2 or T 2** in the submillisecond range, such as bone, 1 , 2 , 3 tendon, 4 , 5 , 6 myelin, 7 , 8 , 9 , 10 lung 11 , 12 , 13 and teeth, 14 , 15 , 16 is receiving increasing attention due to its potential for both clinical diagnosis and basic research. The rapid signal decay of such tissues prevents detection and spatial encoding through conventional echo‐based sequences. Therefore, several dedicated short‐T 2 techniques have been developed, usually avoiding echo formation. 17 One efficient and increasingly used technique is zero echo time (ZTE) imaging, 18 , 19 , 20 , 21 where a frequency‐encoding gradient is switched on before radio‐frequency (RF) excitation and signal is acquired as soon as possible afterwards (Figure 1A). In this way, 3D k‐space is covered with radial center‐out trajectories and spherical support (Figure 1B).
In ZTE sequences, the dead time Δt separating signal excitation and reception prevents acquisition of early data and has a lower limit determined by half the RF pulse duration, transmit‐receive switching and the filter group delay. This leads to a gap in central k‐space, 20 (i.e. no data are available in a sphere of radius kgap centered on the k‐space origin). To avoid related image artifacts, three approaches have been suggested: (a) generating the missing information through algebraic reconstruction 24 ; (b) recovering it through additional acquisitions using Cartesian single‐point imaging (SPI), 25 , 26 as in the pointwise encoding time reduction with radial acquisition (PETRA) technique 21 ; and (c) recovering the data by radial readouts at lower gradient strength, as in the WASPI technique (water‐ and fat‐suppressed proton projection MRI) 19 (Figure 1C).
Algebraic ZTE has the advantages of not requiring additional acquisition and having a benign behavior of the point spread function (PSF) under T 2* decay. However, image reconstruction becomes ill‐conditioned for kgap exceeding three Nyquist dwells (dk) (where dk = 1/FOV and FOV is the field of view), 27 hence preventing application at larger gaps. WASPI retrieves the missing data with a time‐efficient radial acquisition, which, however, leads to a discontinuous T 2 *‐related modulation transfer function (MTF) in k‐space and thus a propensity for increased PSF side lobes and associated oscillatory image artefacts. 22 On the other hand, PETRA is robust against artifacts but hampered by the slow SPI acquisition of the inner k‐space (k < k gap). 22 Therefore, the best sequence choice depends on the particular imaging task and especially on the gap size. However, none of the described methods are well suited for large gaps (i.e. kgap of tens of dwells), because algebraic ZTE and WASPI lead to poor image quality and PETRA causes undesirably long scan times. Yet, imaging under such conditions can be necessary or beneficial. Indeed, large gaps occur at high imaging bandwidth, as required for high‐resolution imaging of short‐T 2 components, in particular in large FOVs, 23 or when the dead time is relatively large, either because of limitations of the RF hardware or by choice to enable T 2 * selection, 21 T 2* mapping, or reduction of chemical‐shift artefacts.
In this work, we explore hybrid filling (HYFI) of the k‐space center as a path to ZTE imaging that is both time‐efficient and robust despite large k‐space gaps. The hybrid approach, which reconciles the advantages of point‐wise and radial filling, was recently sketched in a conference presentation. 28 In the meantime, it has shown promise in a first applied study. 10 , 29 On this basis, the present work has the dual aim of establishing the technical basis of HYFI in due detail and assessing its imaging properties and SNR efficiency in comparison with PETRA. This is done by PSF analysis, 3D imaging simulations, and phantom as well as in vivo imaging.
2. METHODS
2.1. Hybrid filling
The basic idea of HYFI is to replace the SPI part in PETRA by a more time‐efficient acquisition strategy with a pattern that avoids strong discontinuities in the MTF, as occurring in WASPI. This is accomplished by using radial acquisitions, but with a short acquisition time to restrain T 2* decay. To fill the dead‐time gap under this condition, multiple sets of radial acquisitions are used with decreasing gradient strengths. However, around the origin, the encoding speed is so small that the distance traveled within an adequate time is below the distance separating two data points, † dk. In such circumstances, data are acquired single pointwise. This approach results in a hybrid filling of the inner k‐space with a small spherical SPI core surrounded by a stack of shells acquired on radial readouts in an onion‐like fashion (Figure 1C).
To develop an algorithm implementing the principle outlined above, a range R is introduced, to which amplitudes of decaying signals are limited (Figure 2). Dividing R by the signal amplitude at the dead time Δt leads to the amplitude coefficient A describing the allowed range of signal amplitudes in a normalized way. For a given signal decay, selecting a value for A results in tacq, the maximum duration an acquisition may last after Δt. Figure 2 shows that for exponential signal decay as assumed throughout this work follows
(1) |
Based on this parameter, the pattern and timing of the inner k‐space acquisition are derived (i.e. the number and boundaries of concentric subvolumes as well as associated gradient strengths). The gradient strength required to reach a certain k‐value after Δt is
(2) |
whereγ is the gyromagnetic ratio in frequency units. Close to the k‐space origin, small values of G are required for the first points on the grid with spacing dk. With such slow encoding, only a single data point may then fit into tacq. This situation results in a spherical core region where data are acquired single pointwise, and which are acquired efficiently on a Cartesian grid. The radius kSPI of the SPI core is determined by the onset of the radial acquisition when the condition γ G tacq = dk is fulfilled, leading by combination with Equation 2 to
(3) |
At k‐space radii larger than kSPI, at least two points are acquired during tacq. The associated k‐space volumes are concentric shells of thickness γ G tacq, where G is obtained from Equation 2 by setting k to the inner shell radius, starting from kSPI. Note that theoretically kSPI may reach any value between 0 (A = 1) and ∞ (A → 0), but is in practice rounded to multiples of dk and limited to kgap, because it only concerns data located in the inner k‐space.
The HYFI pattern resulting from this procedure is illustrated in Figure 3 in comparison with PETRA and WASPI. Figure 3A shows that in HYFI the inner k‐space is split into subvolumes—the SPI core and a number of shells—such that each of them can be acquired starting after Δt and within tacq. In this way, the amplitudes of exponentially decaying signal are bound to the range R, as shown in the MTF in Figure 3B.
To enable deriving tacq, two parameters need to be selected by the operator: (a) the target T 2* represents the approximate T 2* of the tissues of interest and may be chosen using a rule of thumb that is developed in section 2.2; and (b) the amplitude coefficient A is set to maximally reduce scan time while keeping artifacts to a negligible level. The optimal value depends on the experimental conditions and can be evaluated by means of PSF calculations (as shown below) or by image simulations (as demonstrated in the Results section). By setting A = 0 or A = 1, the HYFI algorithm leads to PETRA and WASPI as limiting cases.
Figure 4 illustrates the tradeoff between scan time reduction and PSF quality in the selection of A. As savings in scan time are largely described by the reduction of the SPI core, Figure 4A shows the number of excitations NSPI required to fill the core as a function of A. In typical cases where Δt is less than T 2*, choosing A ≲ 0.1 is sufficient to decrease the size of the SPI core considerably. Moreover, Figure 4B shows that in such circumstances the PSF lineshape is largely unaffected, suggesting that scan time can indeed be reduced while preserving image quality. More detailed PSFs are shown in the supporting information (Figure S1).
Possible savings in scan time for the complete inner k‐space, including acquisition of both core and shells, are illustrated in Figure 5. To minimize scan time, in each radial shell the angular spoke density is adapted to fulfill the Nyquist criterion at the outer shell radius. In the limiting cases of PETRA and WASPI, the number of RF excitations Ngap required to fill a sphere of radius kgap varies with the volume and the surface of the sphere, respectively. In HYFI, Ngap depends on the amplitude coefficient A and is bounded by these limits. As already suggested in Figure 4, small amplitude coefficients are sufficient to significantly decrease Ngap.
The implementation of the described HYFI algorithm is governed by the discrete nature of sampling k‐space at Nyquist dwell distance. An example implementation is provided as Matlab source code at https://doi.org/10.3929/ethz‐b‐000415045.
2.2. Choice of imaging parameters
The parameters of all imaging experiments are listed in Tables 1 and S8.
TABLE 1.
Sample | Target T 2* | BW (kHz) | FOV (mm) | G (mT/m) | Δr (mm) | M | Δt (μs) | k gap (dk) | NSA | Pulse | Pul. dur (μs) | Coil type | B0 (T) |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Sphere simulations | 100 | 1000 | 200 | 117.4 | 1 | 200 | 30 | 30 | x | x | x | x | x |
Stack of erasers | 100 | 250 | 240 | 24.5 | 1.9 | 128 | 100 | 25 | 1 | Block | 2 | Birdcage | 7 |
Bone | 200 | 766 | 90 | 199.9 | 0.4 | 256 | 40 | 31 | 8 | Block | 2 | Surface | 3 |
Head | 200 | 2000 | 300 | 156.6 | 1.7 | 176 | 15 | 30 | 19 | HSn | 8 | Birdcage | 3 |
Knee | 500 | 250 | 240 | 24.5 | 1.0 | 240 | 5.5 | 2 | 1 | Block | 2 | Birdcage | 7 |
Knee | 500 | 250 | 240 | 24.5 | 1.0 | 240 | 200 | 50 | 1 | Block | 2 | Birdcage | 7 |
MnCl2 vials | 55 | 852 | 100 | 199.9 | 1.0 | 100 | 55 | 47 | 1 | HSn | 10 | Surface | 3 |
MnCl2 vials | 100 | 470 | 100 | 110.4 | 1.0 | 100 | 100 | 47 | 1 | HSn | 10 | Surface | 3 |
MnCl2 vials | 200 | 235 | 100 | 55.2 | 1.0 | 100 | 200 | 47 | 1 | HSn | 10 | Surface | 3 |
MnCl2 vials | 400 | 118 | 100 | 27.6 | 1.0 | 100 | 400 | 47 | 1 | HSn | 10 | Surface | 3 |
MnCl2 vials | 600 | 78 | 100 | 18.3 | 1.0 | 100 | 600 | 47 | 1 | HSn | 10 | Surface | 3 |
In general, small dead times are desired to maximize signal amplitude and reduce scan duration. However, in some situations it may be preferable to deliberately extend the dead time and/or the gap size by inserting an additional delay before data acquisition. For example, short‐T 2 components that cannot be resolved may be selectively suppressed in this way to improve image quality and quantification. Moreover, in the inner k‐space, data samples are acquired in a small time range bounded by the maximum shell acquisition duration t acq. When t acq is small enough, this leads to similar T 2 * weighting and chemical shift‐induced phase for all inner k‐space data. Consequently, increasing the inner k‐space volume to a substantial part of the support improves the accuracy of T 2 * mapping based on a series of such data. In addition, it also reduces PSF blurring 23 as well as chemical shift artifacts. In this work, increased dead time is employed for the suppression of components with extremely short T 2, and reduction of chemical shift artifacts, as well as T 2 * mapping.
When setting up a protocol, values for target T 2* and amplitude coefficient A must be selected. A suitable choice of the parameter pair (A, T 2*) is crucial for optimal HYFI performance. Figure 4 reveals that a good compromise between image quality and scan time is obtained for A ≲ 0.1. Indeed, for target T 2* ≥ Δt, the number of excitations required in the SPI region decreases by more than 95% while PSFs have negligible side lobes. However, in most cases the imaged samples contain multiple signal sources with different relaxation times T 2*. Then a choice of A that is appropriate for a given T 2* leads to stronger decay for signals with shorter T 2* and hence potentially to artifacts. As usually not all relaxation times present in a sample are known a priori, an educated choice of the target T 2* may not always be possible. As a rule of thumb, in the presence of multiple T 2 * it is considered safe to choose a target T 2 * = Δt because signals with T 2 * < Δt will considerably decay before data acquisition. Moreover, as shown in the Results section, the target T 2 * can be chosen to be larger than Δt if the MR signal is dominated by sources with T 2 * >> Δt.
2.3. Hardware
All the experiments were performed on Achieva MRI systems (Philips Healthcare, Best, the Netherlands) at 3 or 7 T, complemented with symmetrically biased transmit‐receive switches 30 with switching times of approximately 3 μs at 3 T and 1 μs at 7 T, custom‐made spectrometers 31 with up to 4 MHz acquisition bandwidth and short digital filters with group delays down to 1.2 μs. Moreover, the 3 T scanner was equipped with a high‐performance gradient insert system capable of reaching 200 mT/m at full duty cycle 32 and a broadband linear RF power amplifier BLA1000‐I E (Bruker Biospin, Wissembourg, France). Largely 1H‐free RF coils were used for both transmission and reception, a surface coil of 80 mm diameter and two birdcage coils. 33 , 34
Block and sweep hyperbolic secant (HSn) pulses 35 with bandwidth matching the imaging bandwidth were used for excitation. HSn pulses were chosen for high bandwidth imaging, where block pulses posed too strong limitations on flip angles due to the combination of limited RF power and short durations. The pulse power was set empirically to obtain maximum signal in the tissues of interest. Flip angles were not calibrated but estimated to range from 2 to 4 degrees.
2.4. Samples
The relaxation constants of the samples were evaluated from mono‐ or double‐exponential fits on free induction decay (FID) signals measured at 3 T.
An imaging phantom with two different materials was created by placing a stack of erasers (Caran d'Ache 0149.340) with T 2 * ≈ 380 μs onto a disk made of rubber with T 2 * ≈ 130 μs.
A bone phantom was taken from a previous study. 22 The piece of bovine tibia of 60 mm diameter and 25 mm thickness had been freed of sources of long‐lived MR signal (i.e. soft tissues). The signal relaxation appeared to be dominated by two T 2 *s of about 10 and 150 μs. Hence, for imaging, a relatively long dead time of 40 μs was chosen deliberately to suppress the shorter component that cannot be resolved to the targeted submillimeter resolution and to focus on the longer‐T 2 contributions, as well as to increase the PSF‐limited resolution. 23
An imaging phantom with a range of T 2 values was created by filling six solutions of MnCl2 with concentrations of 240, 120, 60, 30, 15 and 7.5 mM into glass vials. For measuring the transverse relaxation times, the solutions were filled into glass spheres of 20 mm diameter to minimize susceptibility effects. FID signals were fitted with single exponentials, providing decay constants of 54, 92, 181, 341, 663 and 1271 μs, respectively. For T 2* mapping, a series of images was acquired with constant gap (kgap = 47) and different dead times (Δt = 55, 100, 200, 400 and 600 μs), where the gradient strength was adapted according to Table 1.
In vivo imaging of a knee and a head was conducted in healthy volunteers according to applicable ethics approval, and written informed consent was obtained from all subjects. For knee imaging, the dead time was intentionally increased to 200 μs to extend the inner k‐space and thus reduce chemical shift artifacts.
2.5. Image reconstruction and processing
Images were reconstructed using an iterative conjugate gradient algorithm, 36 which in principle is capable of handling the complex density pattern of HYFI data alone, without the need to introduce specific merging filters. However, for improved convergence, density correction was applied as obtained by an iterative algorithm. 37 Additionally, when modulated HSn pulses were used, RF pulse correction was performed. 38 Finally, geometry and bias field corrections were applied to the head images to compensate for gradient nonlinearity 32 and coil sensitivities.
For determining the SNR in images, additional noise data were acquired in the absence of RF excitation. 39 , 40 The average signal over a region of interest (ROI) in the magnitude sample image was divided by the standard deviation of the noise image in the same ROI. The SNR efficiency was obtained as , assuming the common relation of averaging and SNR. Finally, the relative scan time for equal SNR was calculated according to
(4) |
More details about the SNR analysis are provided in the supporting information.
For interpreting the series of images acquired at different dead time Δt, the echo time is defined as TE = Δt for both PETRA and HYFI, according to the convention that TE indicates the time at which the k‐space center is acquired. T 2* fitting was performed with single‐exponential functions and all amplitudes were normalized with the respective value obtained at t = 0.
2.6. Simulations
Simulations were performed to evaluate artifacts due to T 2* decay in relation to improvements in scan efficiency, as a basis for optimizing the HYFI parameter choice. 1D PSFs were calculated by Fourier transforming MTFs. For 3D simulations, the k‐space signal of spheres with 50, 30, 18 and 9 mm diameter and T 2 of 100 μs was created analytically, 41 assuming identical magnetization after each excitation. Images were reconstructed with the algorithm described above by following the same pipeline as for the experimental data. Corresponding imaging parameters are given in Table 1.
3. RESULTS
The effect of the amplitude coefficient A on image quality and scan efficiency is illustrated with 3D simulations (Figure 6). Corresponding image profiles are shown in Figure S2. As the amplitude coefficient A increases, the relative number of excitations required to fill the gap, n gap, decreases, thus increasing the scan efficiency. For this sample, images without noticeable artifacts are obtained with A less than 0.1.
The phantom experiment in Figure 7 demonstrates the HYFI principle over a large range of A. For A = 0, the inner k‐space is acquired in an SPI fashion, leading there to a constant plateau of T 2 * weighting, maximum n gap and artifact‐free images. When A increases, radial shells with restricted decay replace part of the SPI plateau and n gap diminishes. However, this also creates progressively increasing irregularities in the MTF, which in turn leads to increasingly large ringing artifacts out and inside the imaged object.
In Figure 8, the performance of PETRA and HYFI are compared for imaging a sample of bovine tibia. Both techniques lead to high quality images depicting fine trabecular bone structure. Other details appear in the maximum intensity projection, such as glue from the coil conductor and the fixation tape. In HYFI, the choice of a small A = 0.04 is sufficient to significantly improve the SNR efficiency compared with PETRA, which translates into a 25% decrease of total scan time for the same SNR.
The head images in Figure 9 were acquired with an unusually high bandwidth of 2 MHz and confirm the above results. The SNR efficiency of HYFI is enhanced compared with PETRA and leads to a 40% scan time reduction for the same SNR while preserving image quality.
Figure 10 shows the influence of the dead‐time gap on chemical shift artifacts in ZTE knee imaging. At minimum dead time Δt = 5.5 μs leading to kgap= 1.4 dk, signal intensity is maximized and, moreover, the missing data points can be reconstructed algebraically leading to minimum scan duration. However, due to signal dephasing during the spoke sampling of 480 μs, chemical shift artefacts appear at water‐fat boundaries. 17 Increasing the dead time to 200 μs enlarges the inner k‐space to 50 dk radius and thus reduces the acquisition time range for data located in this region. In PETRA (Figure 10B), signal can dephase only within 1 dk of 4 μs. Hence, while accepting a loss of signal intensity associated with the longer dead time, chemical shift artifacts are strongly reduced and the resolution at water‐fat interfaces is improved. However, scan time is substantially increased. Under the same circumstances, HYFI provides similar image quality but reduces the acquisition duration for the same SNR by 48% compared with PETRA (Figure 10C).
T 2 * mapping of short‐T 2 samples is demonstrated in Figure 11. To enable accurate fitting, the inner k‐space was deliberately increased up to the outer limit of the support such that the whole k‐space was acquired within a restricted time range. In this way, PETRA approaches pure SPI acquisition. 26 Figure 11A shows that at TE = 55 μs, all the vials are well depicted and image intensity drops at larger TE in the short‐T 2 samples. For the same image and T 2 * map quality, the scan time of HYFI is 62% lower than PETRA. Figure 11B shows a good correspondence between the fit of average map intensities and the fit of the FID, especially in the short‐T 2 * range. Mean values including the 95% confidence interval are given in Table S7 in the supporting information. Two observations can be made: (a) there is an increasing divergence between T 2* values fitted from FIDs and images as T 2 * gets larger, and (b) the relaxation times fitted from the HYFI data are slightly but consistently smaller than the PETRA results. The first observation is assigned to residual B0 inhomogeneity in the samples used for FID measurements, leading to smaller effective T 2* values. The second effect results from the fact that the echo time TE is considered equal to the dead time ∆t, which fits PETRA better than HYFI. Indeed, in HYFI, signal is still acquired for a duration t acq after ∆t. During that time, the signal decays and appears smaller than in PETRA. Also, as the target T 2* (and hence t acq) increases with TE (c.f. Table 1), this effect also increases with TE and thus leads to smaller fitted T 2 * values.
4. DISCUSSION
HYFI, a recently introduced ZTE‐based method with hybrid filling of the inner k‐space, was described in detail and its performance in the presence of large k‐space gaps was studied. It was demonstrated that with HYFI, substantial reductions in scan time can be enabled while preserving image quality compared with the PETRA technique. The advantage of HYFI increases with gap size and is therefore of particular interest at high imaging bandwidth, large minimum RF dead times, and large dead times selected to manipulate image contrast. The technique was successfully employed for imaging on different phantoms as well as in vivo and ex vivo.
Using HYFI with high efficiency and fidelity necessitates a suitable choice of the parameter pair (A, T 2*). The simulations in Figure 6 demonstrate that in the presence of a single T 2, A ≲ 0.1 provides considerable savings in scan time at still high image quality. However, as A is increased, artifacts become more likely due to coherent interaction of increasingly large PSF side lobes occurring predominantly in the center of large objects.
In more common cases involving different tissues and molecules and thus a range of transverse relaxation times, the choice of the target T 2* requires extra considerations. Selecting the smallest T 2* present in the sample is safe but usually too conservative. Typically, components with T 2* less than Δt have little influence on the final image and choosing a target T 2* = Δt can be considered an appropriate rule of thumb (Figure 11). Moreover, the target T 2* might be increased to values larger than Δt without degrading image quality if the signal is dominated by components with longer T 2*. In the presented in vivo data, target T 2* of a few hundred microseconds still lead to images without noticeable artifacts (Figures 9 and 10), although components with clearly faster relaxation but lower intensity (e.g. myelin or collagen) are present. Only if the latter signals should be extracted from the data, they need to be considered for setting the target T 2*.
In the performed experiments, the reductions in scan time of HYFI with respect to PETRA range from 25% to 62% (Figures 8, 9, 10, 11). The influencing factors are gap size, spatial resolution and the parameter pair (A, T 2*). The relative number of excitations required to fill the inner k‐space with HYFI decreases with increasing gap size compared with PETRA (Figure 5). This explains why the best HYFI performance is obtained in the examples shown in Figures 10 and 11, where gaps were deliberately increased to high values to reduce chemical shift artefacts and perform T 2* mapping, respectively. The resolution determines the time spent for the acquisition of the outer k‐space, which in turn affects the relative scan time spent on the inner k‐space. Thus, as resolution and hence scan time are increased, the absolute time difference between PETRA and HYFI does not change but the relative advantage of HYFI diminishes. Finally, the selection of the parameter pair (A , T 2*) influences the k‐space trajectory and affects both scan time and image SNR. As the amplitude coefficient A increases, the SPI region decreases and is replaced by radial acquisitions, which leads to higher k‐space data density and thus a reduction of image noise variance. 23 However, the data points experience stronger T 2* weighting (Figure 3), which translates into a smaller integral of the MTF and hence to a smaller PSF main lobe (Figure 4), leading to decreased voxel intensity. These two effects partly compensate each other. For small amplitude coefficients they even largely balance and the improvement of HYFI in SNR efficiency can be well approximated as if arising from scan time reduction alone. High‐resolution imaging of short‐T 2 components benefits from the use of high gradients. 42 As shown in Figure 8, bone tissue with T 2* ≈ 150 μs can be imaged at an isotropic resolution of 400 μm using a gradient strength of 200 mT/m. Such high gradients induce high bandwidths and thus large k‐space gaps, especially in large FOVs as required for imaging in humans (Figure 9). In such circumstances, substituting a large part of the SPI region by radial spokes with HYFI particularly improves scanning efficiency. Moreover, at large gradient amplitude Cartesian SPI acquisition can produce significant mechanical vibrations and acoustic noise due to partly large gradient switching between k‐space directions. With HYFI, as long as the gradient can be used at full duty cycle (i.e. without switching back to zero amplitude), these effects are significantly reduced because k‐space directions are uniformly distributed in all directions and sequentially accessed along spiral spoke ordering 43 requiring slower gradient switching, thus clearly improving patient comfort.
The results of this work indicate the potential of HYFI for direct imaging of ultrashort‐T 2 components such as in the myelin bilayer in the brain. However, as observable in Figure 9, basic ZTE sequences lead to mostly proton density‐weighted images, and some kind of selectivity is required to isolate the tissues of interest. One possibility to achieve T 2* selection uses postprocessing of a series of images acquired after different dead times. An example of such an experiment is shown in Figure 11, where T 2* mapping of MnCl2 solutions was performed by fitting exponential signal decays. This kind of approach has been shown to enable T 2* selection of ultrafast relaxing MR signals in the brain that can potentially be assigned to the myelin bilayer. 10 Further improvements in quantification with HYFI‐based T 2* mapping are expected with a more advanced definition of TE or signal models taking into account the precise sequence timing.
Clinical scanners with state‐of‐the‐art hardware specifications (e.g. an RF switching time of approximately 30 μs, 44 gradient slew rate of approximately 200 mT/m and maximum gradient strength of approximately 80 mT/m, yet at limited duty cycle 17 ) may be used for short‐T 2 PETRA imaging and will lead to gap sizes of about 30–40 dk. Thus, assuming a TR of a few milliseconds, the acquisition of the inner k‐space takes several minutes (c.f. Figure 4). In such a situation and as illustrated in this paper, substantial improvement in SNR efficiency can be expected when using HYFI instead of PETRA. If the same scanners are used with lower bandwidth (e.g. G < 40 mT/m), the advantage of HYFI is limited to lower acoustic noise and reduced mechanical vibrations. In the special case of a combination of low gradients and large dead times (e.g. G = 10 mT/m, RF switching time of approximately 50 μs), the dead time could be used to ramp up the gradient to its target strength, thus allowing smaller excitation bandwidths at reduced gap sizes and avoiding the need for HYFI, yet at the price of reduced spatial resolution. 23 A similar idea was exposed in the ramped hybrid encoding (RHE) technique, 45 where the readout gradient is lowered during RF excitation and increased to full strength afterwards during data acquisition. In such cases, k‐space calibration is required because timing errors and eddy current effects distort the k‐space trajectory. Other alternatives to PETRA and hence HYFI are SWIFT 46 and cSWIFT, 47 where gaps are very small or inexistent, respectively. However, the first one comes at an SNR penalty and limited bandwidth 48 and the second approach is particularly sensitive to RF coil loading variations.
Finally, there are situations where a large part of the data should be acquired within a small time range as required for chemical shift artifact reduction or T 2* mapping. For example, for performing the latter, an SPI acquisition of the whole k‐space support is favorable, as shown in Figure 11. Note that in this context, gradients should be switched on before the RF pulse because large gaps are actually targeted. Hence, even with clinical scanners, HYFI can be considered an efficient alternative to SPI methods 26 , 49 in situations where creating gradient echoes is hampered by limited gradient performance.
5. CONCLUSION
The HYFI technique provides both high SNR efficiency and image quality, thus outperforming previously known ZTE‐based pulse sequences. It is particularly advantageous in situations involving large dead times or high gradient strengths where PETRA suffers from long and noisy SPI acquisitions. Promising applications include direct imaging of ultrashort T 2 components, such as the myelin bilayer or collagen, T 2* mapping of ultrafast relaxing signals, and ZTE imaging with reduced chemical shift artifacts.
Supporting information
Froidevaux R, Weiger M, Rösler MB, Brunner DO, Pruessmann KP. HYFI: Hybrid filling of the dead‐time gap for faster zero echo time imaging. NMR in Biomedicine. 2021;34:e4493. 10.1002/nbm.4493
The copyright line for this article was changed on 16 March 2021 after original online publication.
ENDNOTES
In the tissues targeted in this work, short T 2 largely governs T 2 *. Therefore, to describe tissue properties and signal relaxation in general, the term T 2 is used, whereas T 2 * is employed to characterize the decay in actual signal acquisition.
Throughout this manuscript, the term “data points” corresponds to k‐space samples separated by Nyquist distance, dk = 1/FOV.
The exact number of points cannot be calculated analytically. It corresponds to the number of nodes of a Cartesian grid which have a radial distance to the k‐space center smaller than kgap. However, a very good approximation can be found by multiplying the volume of the gap (given by the above formula) with the density of data points acquired on the Cartesian grid, (which in this case is 1 data point per dk3). In order to obtain an integer number, kgap may be rounded to the closest integer value.
DATA AVAILABILITY STATEMENT
The Matlab files that support the findings of this study are openly available in the ETH research collection in “Supporting material for HYFI paper” at https://doi.org/10.3929/ethz‐b‐000415045.
REFERENCES
- 1. Cao H, Nazarian A, Ackerman JL, et al. Quantitative 31P NMR spectroscopy and 1H MRI measurements of bone mineral and matrix density differentiate metabolic bone diseases in rat models. Bone. 2010;46(6):1582‐1590. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Du J, Bydder GM. Qualitative and quantitative ultrashort‐TE MRI of cortical bone. NMR Biomed. 2013;26(5):489‐506. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Wehrli FW. Magnetic resonance of calcified tissues. J Magn Reson. 2013;229:35‐48. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Peto S, Gillis P, Henri VP. Structure and dynamics of water in tendon from NMR relaxation measurements. Biophys J. 1990;57(1):71‐84. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Robson MD, Benjamin M, Gishen P, Bydder GM. Magnetic resonance imaging of the Achilles tendon using ultrashort TE (UTE) pulse sequences. Clin Radiol. 2004;59(8):727‐735. [DOI] [PubMed] [Google Scholar]
- 6. Rahmer J, Börnert P, Groen J, Bos C. Three‐dimensional radial ultrashort echo‐time imaging with T2 adapted sampling. Magn Reson Med. 2006;55(5):1075‐1082. [DOI] [PubMed] [Google Scholar]
- 7. Nayak KS, Pauly JM, Gold GE, Nishimura DG. Imaging ultra‐short T2 species in the brain. Proc., ISMRM, 8th Annual Meeting, Denver 2000:509.
- 8. Du J, Ma G, Li S, et al. Ultrashort echo time (UTE) magnetic resonance imaging of the short T2 components in white matter of the brain using a clinical 3T scanner. Neuroimage. 2014;87:32‐41. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Wilhelm MJ, Ong HH, Wehrli SL, et al. Direct magnetic resonance detection of myelin and prospects for quantitative imaging of myelin density. Proc Natl Acad Sci U S A. 2012;109(24):9605‐9610. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Weiger M, Froidevaux R, Baadsvik EL, Brunner DO, Rösler MB, Pruessmann KP. Advances in MRI of the myelin bilayer. Neuroimage. 2020;217:116888. [DOI] [PubMed] [Google Scholar]
- 11. Bergin CJ, Pauly JM, Macovski A. Lung parenchyma: Projection reconstruction MR imaging. Radiology. 1991;179(3):777‐781. [DOI] [PubMed] [Google Scholar]
- 12. Kuethe DO, Adolphi NL, Fukushima E. Short data‐acquisition times improve projection images of lung tissue. Magn Reson Med. 2007;57(6):1058‐1064. [DOI] [PubMed] [Google Scholar]
- 13. Gibiino F, Sacolick L, Menini A, Landini L, Wiesinger F. Free‐breathing, zero‐TE MR lung imaging. Magn Reson Mater Physics Biol Med. 2015;28(3):207‐215. [DOI] [PubMed] [Google Scholar]
- 14. Gruwel MLH, Latta P, Tanasiewicz M, Volotovskyy V, Šramek M, Tomanek B. MR imaging of teeth using a silent single point imaging technique. Appl Phys a Mater Sci Process. 2007;88(4):763‐767. [Google Scholar]
- 15. Hövener JB, Zwick S, Leupold J, et al. Dental MRI: Imaging of soft and solid components without ionizing radiation. J Magn Reson Imaging. 2012;36(4):841‐846. [DOI] [PubMed] [Google Scholar]
- 16. Weiger M, Pruessmann KP, Bracher AK, et al. High‐resolution ZTE imaging of human teeth. NMR Biomed. 2012;25(10):1144‐1151. [DOI] [PubMed] [Google Scholar]
- 17. Weiger M, Pruessmann KP. Short‐T2 MRI: Principles and recent advances. Prog Nucl Magn Reson Spectrosc. 2019;115:237‐270. [DOI] [PubMed] [Google Scholar]
- 18. Hafner S. Fast imaging in liquid and solids with the back‐projection low angle shot (blast) technique. Magn Reson Imaging. 1994;12(7):1047‐1051. [DOI] [PubMed] [Google Scholar]
- 19. Wu Y, Dai G, Ackerman JL, et al. Water‐ and fat‐suppressed proton projection MRI (WASPI) of rat femur bone. Magn Reson Med. 2007;57:554‐567. [DOI] [PubMed] [Google Scholar]
- 20. Weiger M, Pruessmann KP. MRI with zero echo time. Encycl Magn Reson. 2012;1:311‐322. [Google Scholar]
- 21. Grodzki DM, Jakob PM, Heismann B. Ultrashort echo time imaging using pointwise encoding time reduction with radial acquisition (PETRA). Magn Reson Med. 2012;67(2):510‐518. [DOI] [PubMed] [Google Scholar]
- 22. Froidevaux R, Weiger M, Brunner DO, Dietrich BE, Wilm BJ, Pruessmann KP. Filling the dead‐time gap in zero echo time MRI: Principles compared. Magn Reson Med. 2018;79(4):2036‐2045. [DOI] [PubMed] [Google Scholar]
- 23. Froidevaux R, Weiger M, Rösler MB, et al. High‐resolution short‐T2 MRI using a high‐performance gradient. Magn Reson Med. 2020;84(4):1933‐1946. [DOI] [PubMed] [Google Scholar]
- 24. Kuethe DO, Caprihan A, Lowe IJ, Madio DP, Gach HM. Transforming NMR data despite missing points. J Magn Reson. 1999;139(1):18‐25. [DOI] [PubMed] [Google Scholar]
- 25. Emid S. Ultra high resolution multiple quantum spectroscopy in solids. Phys Ther. 1985;128B:79‐80. [Google Scholar]
- 26. Balcom BJ, Macgregor RP, Beyea SD, Green DP, Armstrong RL, Bremner TW. Single‐point ramped imaging with T1 enhancement (SPRITE). J Magn Reson. 1996;123(1):131‐134. [DOI] [PubMed] [Google Scholar]
- 27. Weiger M, Brunner DO, Tabbert M, Pavan M, Schmid T, Pruessmann KP. Exploring the bandwidth limits of ZTE imaging: Spatial response, out‐of‐band signals, and noise propagation. Magn Reson Med. 2015;74(5):1236‐1247. [DOI] [PubMed] [Google Scholar]
- 28. Froidevaux R, Weiger M, Rösler MB, Brunner DO, Pruessmann KP. HYFI: Hybrid filling of the dead‐time gap for faster zero echo time imaging. Proc. 27th Annual Meeting ISMRM, Montreal, Canada. 2019:943. [DOI] [PMC free article] [PubMed]
- 29. Baadsvik EL, Weiger M, Froidevaux R, et al. High‐resolution MRI of mummified tissues using advanced short‐T2 methodology and hardware. Magn Reson Med. 2020;85(3):1481‐1492. [DOI] [PubMed] [Google Scholar]
- 30. Brunner DO, Furrer L, Weiger M, et al. Symmetrically biased T/R switches for NMR and MRI with microsecond dead time. J Magn Reson. 2016;263:147‐155. [DOI] [PubMed] [Google Scholar]
- 31. Dietrich BE, Brunner DO, Wilm BJ, et al. A field camera for MR sequence monitoring and system analysis. Magn Reson Med. 2016;75(4):1831‐1840. [DOI] [PubMed] [Google Scholar]
- 32. Weiger M, Overweg J, Rösler MB, et al. A high‐performance gradient insert for rapid and short‐T2 imaging at full duty cycle. Magn Reson Med. 2018;79(6):3256‐3266. [DOI] [PubMed] [Google Scholar]
- 33. Weiger M, Brunner DO, Schmid T, et al. A virtually 1H‐free birdcage coil for zero echo time MRI without background signal. Magn Reson Med. 2017;78(1):399‐407. [DOI] [PubMed] [Google Scholar]
- 34. Rösler MB, Weiger M, Brunner DO, Schmid T, Pruessmann KP. An RF birdcage coil designed for an insert gradient coil dedicated to short‐T2 MRI. Proc. 26th Annual Meeting ISMRM, Honolulu. 2017:2668.
- 35. Schieban K, Weiger M, Hennel F, Boss A, Pruessmann KP. ZTE imaging with enhanced flip angle using modulated excitation. Magn Reson Med. 2014;74:684‐693. [DOI] [PubMed] [Google Scholar]
- 36. Pruessmann KP, Weiger M, Börnert P, Boesiger P. Advances in sensitivity encoding with arbitrary k‐space trajectories. Magn Reson Med. 2001;46(4):638‐651. [DOI] [PubMed] [Google Scholar]
- 37. Zwart NR, Johnson KO, Pipe JG. Efficient sample density estimation by combining gridding and an optimized kernel. Magn Reson Med. 2012;67(3):701‐710. [DOI] [PubMed] [Google Scholar]
- 38. Li C, Magland JF, Seifert AC, Wehrli FW. Correction of excitation profile in zero echo time (ZTE) imaging using quadratic phase‐modulated RF pulse excitation and iterative reconstruction. IEEE Trans Med Imaging. 2014;33(4):961‐969. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Dietrich O, Raya JG, Reeder SB, Reiser MF, Schoenberg SO. Measurement of signal‐to‐noise ratios in MR images: Influence of multichannel coils, parallel imaging, and reconstruction filters. J Magn Reson Imaging. 2007;26(2):375‐385. [DOI] [PubMed] [Google Scholar]
- 40. NEMA . NEMA Standards Publication MS 1–2001: Determination of Signal‐to‐Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging. Natl Electr Manuf Assoc. 2008;2014:1‐37. [Google Scholar]
- 41. Cheng GK, Sarlls JE, Özarslan E. Three‐dimensional analytical magnetic resonance imaging phantom in the Fourier domain. Magn Reson Med. 2007;58:430‐436. [DOI] [PubMed] [Google Scholar]
- 42. Froidevaux R, Weiger M, Rösler MB, et al Ultra‐high‐bandwidth, high‐resolution MRI of fast relaxing spins. Proc. 26th Annual Meeting ISMRM, Honolulu. 2017:4037.
- 43. Wong STS, Roos MS. A strategy for sampling on a sphere applied to 3D selective RF pulse design. Magn Reson Med. 1994;32(6):778‐784. [DOI] [PubMed] [Google Scholar]
- 44. Wiesinger F, Sacolick LI, Menini A, et al. Zero TE MR bone imaging in the head. Magn Reson Med. 2015;75:107‐114. [DOI] [PubMed] [Google Scholar]
- 45. Jang H, Wiens CN, McMillan AB. Ramped hybrid encoding for improved ultrashort echo time imaging. Magn Reson Med. 2016;76(3):814‐825. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46. Garwood M, Idiyatullin D, Corum CA, et al. Capturing signals from fast‐relaxing spins with frequency‐swept MRI: SWIFT. Encycl Magn Reson. 2012;1:367‐380. [Google Scholar]
- 47. Idiyatullin D, Suddarth S, Corum CA, Adriany G, Garwood M. Continuous SWIFT. J Magn Reson. 2012;220:26‐31. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48. Weiger M, Pruessmann KP, Hennel F. MRI with zero echo time: Hard versus sweep pulse excitation. Magn Reson Med. 2011;66(2):379‐389. [DOI] [PubMed] [Google Scholar]
- 49. Kaffanke JB, Romanzetti S, Dierkes T, Leach MO, Balcom BJ, Shah NJ. Multi‐Frame SPRITE: A method for resolution enhancement of multiple‐point SPRITE data. J Magn Reson. 2013;230:111‐116. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The Matlab files that support the findings of this study are openly available in the ETH research collection in “Supporting material for HYFI paper” at https://doi.org/10.3929/ethz‐b‐000415045.