Table 2.

Differences between Systolic and Diastolic BP Measured Using Intra-Arterial, Auscultatory, and Oscillometric Methods

Systolic Mean ± SD Diastolic Mean ± SD
Intra-arterial method 125.47 ± 21.39 64.59 ± 11.18
Auscultatory method 120.68 ± 19.57 66.07 ± 10.26
Difference 4.79 ± 5.54 −1.47 ± 4.71
p value* t = 11.59 p = .000 t = −4.20. p = .000
Intra-arterial method 125.47 ± 21.39 64.59 ± 11.18
Oscillometric method 121.13 ± 19.04 73.91 ± 10.62
Difference 4.33 ± 12.11 −9.31 ± 7.98
p value* t = 4.80. p = .000 t = −15.66. p = .000
Auscultatory 120.68 ± 19.57 66.07 ± 10.26
Oscillometric 121.13 ± 19.04 73.91 ± 10.62
Difference −1.55 ± 9.52 −7.84 ± 7.38
p value* t = −.64. p = .552 t = −14.24. p = .000

Note.

*

Paired samples t test.