Skip to main content
. 2021 Jun 17;8:661064. doi: 10.3389/fsurg.2021.661064

Table 2.

Extrahepatic bile duct (EBD) sealing by two acrosurg devices.

Outcome AS AT p*
Number of trials 6 6
Diameter (mm) 1.8 [1.8–1.8] 3.4 [2.8–3.4] 0.0045
Burst pressure (mmHg) 607.5 [472–705] 806.5 [776–889] 0.0161
*

p-value from Wilcoxon's ranksum test, data represented as median [interquartile range].

AS, Acrosurg Scissors; AT, Acrosurg Tweezers.