Table 2.

Variation in image noise and CNR in the hepatic parenchyma with respect to BMI category

Parameter Normal BMI (n = 45) Pre-obese (n = 33) Obese (n = 11) Obese vs normal
p value
Obese vs pre-obese
p value
Image noise
 FBP 22.5 ± 1.9 23.2 ± 1.5 25.6 ± 5.5 0.001 0.01
 FIRST 14.3 ± 1.5 14.9 ± 1.5 15.8 ± 1.5 0.008 0.16
 AiCE 9.9 ± 0.8 10.0 ± 0.8 10.4 ± 1.2 0.13 0.18
CNR
 FBP 2.2 ± 0.7 2.2 ± 0.8 1.4 ± 0.8 0.002 0.002
 FIRST 3.1 ± 0.9 3.0 ± 1.1 1.7 ± 1.3 <0.001 0.001
 AiCE 4.3 ± 1.2 4.0 ± 1.6 2.4 ± 2.0 <0.001 0.006

AiCE, Advanced intelligent clear-IQ engine; BMI, body mass index; CNR, contrast-to-noise ratio; FBP, filtered back projection; FIRST, Forward projected model-based Iterative reconstruction solution.

Data are presented as mean values ± standard deviation. Dunnett’s test was used for statistical comparisons. The mean volume CT dose index was 5.4 mGy, 9.4 mGy, and 16.5 mGy for the normal BMI, pre-obese, and obese groups, respectively.