Skip to main content
. 2021 Jul;11(7):3051–3058. doi: 10.21037/qims-20-1158

Table 3. Results of qualitative scores and quantitative measurements.

50% ASIR-V 100% ASIR-V DLIR-H Statistical value P value
Subjective evaluation#
   Overall noise 2.27±0.55 3.36±0.58 4.05±0.21 76.91 <0.001
   Vascular margin 3.77±0.61* 2.86±0.56 4.05±0.58* 24.79 <0.001
   Vascular contrast 3.14±0.64 4.00±0.62* 4.14±0.64* 16.19 <0.05
Objective evaluation&
   Aorta CT 412.50±118.05* 413.00±118.00* 413.41±118.18* 0.00 1.00
   Aorta SD 29.45±7.59 20.45±6.93* 19.24±5.77* 14.78 <0.001
   Aorta SNR 14.64±4.66 21.53±6.8* 22.58±7.04* 10.45 <0.001
   Aorta CNR 13.73±4.00 21.6±5.76* 20.68±6.00* 14.35 <0.001

SD, image noise value. *, without in-group statistical difference; #, subjective scores were compared with Friedman’s test; &, objective scores were compared with one-way ANOVA. ASIR-V, adaptive statistical iterative reconstruction-V; DLIR-H, high-setting deep learning image reconstruction; SNR, signal-to-noise ratio; CNR, contrast-to-noise ratio.