Table II.

Gamma passing rates for pixels >10% prescribed dose comparing motion management strategies for periodic motion with and without added baseline drift to a static reference case. None refers to a scenario where the tracking plan was delivered, but no actual tracking was applied.

Periodic motion Periodic+drift motion
1%/1mm 2%/2mm 3%/3mm 1%/1mm 2%/2mm 3%/3mm
Central delivery
Tracking
none 23 44 58 14 29 44
w/o predictor 59 83 97
w/ predictor 96 100 100 93 100 100
Mid‐position
conventional 15 29 48
trailing 98 99 100
Peripheral delivery
Tracking
none 22 44 58
w/o predictor 40 75 97
w/ predictor 93 100 100 88 99 100
Mid‐position
conventional 8 15 22
trailing 98 100 100