Table 3. Concordance correlation coefficient, Pearson (precision) and accuracy (bias correction factor) of the X1 PRO and HT IVOS CASA devices in evaluating sperm concentration, TM, and PR compared to MSA.

Device Concordance correlation coefficient Pearson (ρ) Bias correction factor (Cb)
Concentration TM PR Concentration TM PR Concentration TM PR
X1 PRO (n=101) 0.96 0.93 0.91 0.96 0.94 0.92 0.99 0.99 0.99
HT IVOS CASA (n=100) 0.97 0.89 0.92 0.97 0.92 0.93 0.99 0.97 0.99

PR: progressive motility, TM: total motility, MSA: manual semen analysis.